form
prepacked column
crosslinking
cross-linked
packaging
pack of 1 × 5 mL
manufacturer/tradename
Cytiva 17-5080-01
storage condition
(20% Ethanol)
parameter
0.5 Mpa max. pressure, 20 mL/min max. flow rate, 5 mL/min flow rate
technique(s)
liquid chromatography (LC): suitable
bed volume
5 mL
column I.D. × L
16 mm × 25 mm
matrix
4% cross-linked agarose
matrix active group
recombinant protein A (E. coli)
particle size
~90 μm
cleaning
3-12
operating pH
3-10
capacity
~35 mg binding capacity (human lgG / ml medium)
separation technique
affinity
storage temp.
2-8°C
General description
Capacity of HiTrap rProtein A for some monoclonal antibodies. Running conditions: Binding buffer: 20 mM sodium phosphate (incl. 3 M NaCl for IgG), pH 7.0. Elution buffer: 0.1 M sodium citrate, pH 3.0.
HiTrap® rProtein A FF columns are prepacked with rProtein A Sepharose™ Fast Flow for purification and fractionation of IgG subclasses and IgG from different sources, including cell supernatants and ascites.
The recombinant protein A ligand (produced in E. coli) is coupled to Sepharose™ Fast Flow by a technique which generates a stable thioether linkage between rProtein A and the base matrix. The coupling technique is optimized to give high binding capacity for IgG.
The recombinant protein A ligand (produced in E. coli) is coupled to Sepharose™ Fast Flow by a technique which generates a stable thioether linkage between rProtein A and the base matrix. The coupling technique is optimized to give high binding capacity for IgG.
pH below 3 is sometimes required to elute strongly bound IgG species. However, protein ligands may hydrolyze at very low pH.
Features and Benefits
- Quick, convenient preparative purification of monoclonal antibodies.
- Well-established protein A-based medium used in many approved MAb processes.
- Recombinant protein A exhibits similar Fc region specificity to that of native protein A but shows enhanced binding capacity and fewer regulatory concerns due to the total absence of mammalian culture in the ligand production and purification.
- Simple operations with a syringe, pump, or high-performance chromatography system, such as AKTA design.
Preparation Note
Please be aware this product may be shipped 90 days before the expiration date. For more information on the batch specific expiration date, please contact technical service.
Analysis Note
To view the Certificate of Analysis for this product, please visit www.cytiva.com.
Other Notes
The pressure over the packed bed varies depending on a range of parameters such as the characteristics of the chromatography medium and the column tubing used.