70006-U

BIOshell A120 Oligo C18 (2.7 µm) HPLC Column

new

L × I.D. 5 cm × 3.0 mm, pk of 1 ea

Sign Into View Organizational & Contract Pricing

Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist
Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist

Agency

suitable for USP L1

Quality Level

feature

endcapped

packaging

pk of 1 ea

extent of labeling

5.6% Carbon loading

parameter

85 °C temp. limit (at pH 9)
90 °C temp. limit (at pH 2)

L × I.D.

5 cm × 3.0 mm

surface area

75 m2/g

matrix

superficially porous particle particle platform

matrix active group

C18 phase
dimethyloctadecylsilane, surface modified phase

particle size

2.7 μm

pore size

120 Å pore size

pH range

2-9

Looking for similar products? Visit Product Comparison Guide

Related Categories

General description

BIOshell A120 OLIGO C18 is a high-performance liquid chromatography column based on the Fused-Core® particle design, enabling high throughput applications. This particle design exhibits very high column efficiency due to the shallow diffusion paths in the 0.5 µm deep porous shell and the small overall particle size of 2.7 µm.

BIOshell A120 OLIGO C18 has a surface modification that leads to alkaline resistance that provides a stable, reversed-phase packing that can be used for separations of oligonucleotides up to 60 bases at high pH mobile phaseconditions.

BIOshell A120 OLIGO C18 uses surface passivated column hardware to eliminate non-specific adsorption.

Application

The BIOshell A120 OLIGO C18 bonded phase is a dimethyloctadecylsilane that has been specifically designed for applications requiring mobile phases that are mixtures of methanol and water or acetonitrile and water with mobile phase buffers or additives such as HFIP/TEA and TEAA. Using elevated temperatures (e.g., 40 – 60 ºC) will reduce the viscosityof the mobile phase and allow the use of faster flow rates and lower column pressure for high sample throughput. Gradient elution may start at 5% organic and then increase to a higher organic concentration depending on the length and sequence of the oligomers being analyzed. BIOshellA120 OLIGO C18 columns are highly suited for ion pair reversed-phase separationof oligonucleotides up to 60 bases. Additional information on solvent selection and separation techniques can be found in Chapter 13, Practical HPLC MethodDevelopment, Second Edition, L.R. Snyder, J.L. Glajch, and J.J. Kirkland, (JohnWiley & Sons, 1997). The use of MS detectors may require additional considerations for ionization efficiencies and ionization suppression for choices of mobile phase additives. For more information, see Mobile Phase Aging and Its Impact on Electrospray Ionization of Oligonucleotides in J. Am. Soc. MassSpectrom. 2023, 34, 12, 2691–2699. (https://doi.org/10.1021/jasms.3c00264).

Features and Benefits

  • 120 Å pore size, enables separations of oligomers up to 60 bases in length
  • High pH and temperature stability, designed for conditions suited for oligonucleotide separations
  • UHPLC and mass spectrometry compatible stationary phase
  • Surface passivated column hardware to reduce potential of stainless steel sample adsorption

Preparation Note

To maximize column life, ensure that samples and mobile phases are particle-free. The use of guard columns or an in-line filter with 0.5 µm porosity between the sample injector and the column is highly recommended. Should the operating pressure of the column suddenly increase beyond normal levels, reversing the flow direction of the column may be attempted to remove debris on the inlet frit. Long-term storage of silica-based, reversed-phase columns is best in 100% acetonitrile. Columns maybe safely stored for short periods (up to 3 or 4 days) in most common mobile phases. However, when using buffers, it is best to protect both the column and the HPLC equipment and remove the salts by flushing the column with the same mobile phase without the buffer (e.g., when using 60/40 MeCN/buffer, flush the column with 60/40 MeCN/H2O) to eliminate any danger from corrosion from the salts while providing rapid re-equilibration of the column with the original mobile phase. Before storing the column, the end-fittings should be tightly sealed with the end-plugs that came with the column to prevent the packing from drying.

Other Notes

  1. A new column contains a mixture of water and acetonitrile. Initial care should be taken to avoid mobile phases that are immiscible with this solvent or could cause a precipitate. Water and all common organic solvents are compatible with BIOshell A120 OLIGO C18 columns.
  2. BIOshell A120 OLIGO C18 columns are rated to pH 9 at 85 °C and pH 2 at 90 °C.
  3. Column lifetime may be reduced if the column is continuously run at the maximum temperature and maximum pH.
  4. BIOshell A120 OLIGO C18 columns are stable to operating pressures up to 600 bar (9000 psi).

Legal Information

Fused-Core is a registered trademark of Advanced Materials Technology, Inc.

Regulatory Information

新产品

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Don't see the Right Version?

If you require a particular version, you can look up a specific certificate by the Lot or Batch number.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service