HPLC Analysis of Nitrosamines on Supel Carbon LC

HPLC Analysis of Nitrosamines on Supel™ Carbon LC

Materials

analytical column

Product No.
Description
Pricing

Supel Carbon LC, 2.7 μm HPLC Column

column L × I.D. 10 cm × 3.0 mm

mobile phase component

Product No.
Description
Pricing

Trifluoroacetic acid

eluent additive for LC-MS, LiChropur, ≥99.0% (GC)

Acetonitrile

gradient grade for liquid chromatography LiChrosolv® Reag. Ph Eur

standard

Product No.
Description
Pricing

N-Nitrosodiethylamine (NDEA)

Pharmaceutical Secondary Standard; Certified Reference Material

N-Nitrosodiisopropylamine (NDIPA)

certified reference material, pharmaceutical secondary standard

N-Nitrosodibutylamine (NDBA)

certified reference material, pharmaceutical secondary standard

N-Nitrosoethylisopropylamine (NEIPA / NIPEA)

certified reference material, pharmaceutical secondary standard

N-Nitrosomethylaminobutyric Acid (NMBA)

certified reference material, pharmaceutical secondary standard

N-Nitrosomethylphenylamine (NMPA)

certified reference material, pharmaceutical secondary standard

water purification system

Product No.
Description
Pricing

Milli-Q® IQ 7005 Ultrapure and Pure Water Purification System

Produces ultrapure (Type 1) water and pure (Type 2) water with a production flow rate of 5 L/hr from tap water feed.

CONDITIONS

column

Supel Carbon LC 2.7 µm 100 x 3.0 mm I.D. (59993-U)

mobile phase

[A] Water + 0.1% TFA,[B] Acetonitrile + 0.1% TFA

gradient

0-0.5 min 2.5% B, in 4.5 min to 95% B, 2 min 95% B, 3 min at 2.5% B

flow rate

0.5 mL/min

pressure

80 – 130 bar

column temp.

90 °C

detector

UV, 230 nm

injection

5 µL

sample/matrix

six nitrosamines standard mix (~20 mg/L each) in mobile phase A (water + 0.1% TFA)

Description

General description

Nitrosamines have been known to exist in the environment and within food sources, with awareness further raised in 2018 when many regulatory authorities for medicines became aware of the dangers of N-Nitrosamines as drug contaminants. This class of compounds is known to be potent carcinogens and can be introduced through multiple introduction pathways. In this application, 2.7 µm porous graphitic carbon particles in Supel Carbon LC column are used to separate six different nitrosamines, followed by UV detection.

Analysis Note

The Supel Carbon LC column is used for the separation of six nitrosamine compounds. The column is able to separate six nitrosamines based on reversed phase with good reproducibility for repeated analyses (data not shown here). The resolution and peak asymmetry for the six different nitrosamines are greater than 1.5 and less than 2.0, respectively. The LOD and LOQ of the tested nitrosamines are in the low mg/L range and can be further improved with the use of mass spectrometry instrumentation.

Other Notes

APP_661