HPLC Analysis of Barbiturates in Serum on Discovery® C18 after SPE using Discovery® DSC-18Lt

HPLC Analysis of Barbiturates in Serum on Discovery® C18 after SPE using Discovery® DSC-18Lt application for SPE, application for HPLC

材料

SPE 管或板

产品编号
说明
价格

Discovery® DSC-18Lt 固相萃取管

bed wt. 500 mg, volume 3 mL, pk of 54

分析色谱柱

产品编号
说明
价格

Discovery® C18 高效液相色谱柱

5 μm particle size, L × I.D. 15 cm × 4.6 mm

标准

产品编号
说明
价格

司可巴比妥标准液 溶液

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

苯巴比妥标准液 溶液

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

仲丁巴比妥标准液 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

5种巴比妥混标 溶液

250 μg/mL each component in methanol, ampule of 1.0 mL, certified reference material, Cerilliant®

戊巴比妥标准液 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

甲苯巴比妥

United States Pharmacopeia (USP) Reference Standard

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

0.5 mL porcine serum spiked with 0.5 μg/mL or 1.0 μg/mL each analyte then diluted with 0.5 mL water

SPE tube/cartridge

Discovery DSC-18Lt, 500 mg/3 mL (52613-U)

condition

2 mL methanol; 2 mL DI water

sample addition

1 mL at 0.75 mL/min

washing

2 mL 5% methanol, then vacuum or air dry for 5-10 min

elution

1-2 mL methanol

eluate post-treatment

dry eluate with nitrogen purge (40 °C; 15-20 min), reconstitute in 20 μL mobile phase

column

Discovery C18, 15 cm × 4.6 mm, 5 μm preceded by a 2 cm C18 guard column and 0.5 μm frit filter (504955)

mobile phase

[A] methanol; [B] water (40:60, A:B)

flow rate

1 mL/min

column temp.

30 °C

detector

UV, 214 nm

injection

30 μL, diluted porcine serum extract

说明

分析说明

Barbiturates are commonly abused and among the most widely tested compounds in clinical, forensic, or therapeutic drug monitoring applications. Shown here is the baseline separation of a set of barbiturates on a Discovery C18 HPLC column after extraction from spikd serum with Discovery DSC-18Lt SPE. Highest grade HPLC solvents were used to supply low background interference and low particulate contamination for robust, trouble-free operation. Cerilliant and Sigma-Aldrich reference standards provided reliable identification and quantification. A Zymark® RapidTrace® SPE Workstation was used in this study.