HPLC Analysis of Nitrosamines on Supel Carbon LC

HPLC Analysis of Nitrosamines on Supel™ Carbon LC

材料

water purification system

产品编号
说明
价格

Milli-Q® IQ 7005 超纯和纯水净化系统

Produces ultrapure (Type 1) water and pure (Type 2) water with a production flow rate of 5 L/hr from tap water feed.

分析色谱柱

产品编号
说明
价格

Supel Carbon LC, 2.7 μm HPLC Column

column L × I.D. 10 cm × 3.0 mm

标准

产品编号
说明
价格

N-亚硝基二乙胺 (NDEA)

Pharmaceutical Secondary Standard; Certified Reference Material

N-Nitrosodiisopropylamine (NDIPA)

certified reference material, pharmaceutical secondary standard

N-亚硝基二丁胺 (NDBA)

certified reference material, pharmaceutical secondary standard

N-Nitrosoethylisopropylamine (NEIPA / NIPEA)

certified reference material, pharmaceutical secondary standard

N-Nitrosomethylaminobutyric Acid (NMBA)

certified reference material, pharmaceutical secondary standard

N-Nitrosomethylphenylamine (NMPA)

certified reference material, pharmaceutical secondary standard

流动相组分

产品编号
说明
价格

三氟乙酸

eluent additive for LC-MS, LiChropur, ≥99.0% (GC)

乙腈

gradient grade for liquid chromatography LiChrosolv® Reag. Ph Eur

CONDITIONS

column

Supel Carbon LC 2.7 µm 100 x 3.0 mm I.D. (59993-U)

mobile phase

[A] Water + 0.1% TFA,[B] Acetonitrile + 0.1% TFA

gradient

0-0.5 min 2.5% B, in 4.5 min to 95% B, 2 min 95% B, 3 min at 2.5% B

flow rate

0.5 mL/min

pressure

80 – 130 bar

column temp.

90 °C

detector

UV, 230 nm

injection

5 µL

sample/matrix

six nitrosamines standard mix (~20 mg/L each) in mobile phase A (water + 0.1% TFA)

说明

一般描述

Nitrosamines have been known to exist in the environment and within food sources, with awareness further raised in 2018 when many regulatory authorities for medicines became aware of the dangers of N-Nitrosamines as drug contaminants. This class of compounds is known to be potent carcinogens and can be introduced through multiple introduction pathways. In this application, 2.7 µm porous graphitic carbon particles in Supel Carbon LC column are used to separate six different nitrosamines, followed by UV detection.

分析说明

The Supel Carbon LC column is used for the separation of six nitrosamine compounds. The column is able to separate six nitrosamines based on reversed phase with good reproducibility for repeated analyses (data not shown here). The resolution and peak asymmetry for the six different nitrosamines are greater than 1.5 and less than 2.0, respectively. The LOD and LOQ of the tested nitrosamines are in the low mg/L range and can be further improved with the use of mass spectrometry instrumentation.

其他说明

APP_661