Skip to Content
Merck
CN

MSP010012

Millipak® Barrier Capsule Media: Durapore® 0.22 µm

Chemistry: Hydrophobic/Hydrophilic Polyvinylidene Fluoride (PVDF) Inlet: 9/16 in. Hose Barb Outlet: 9/16 in. Hose Barb

Synonym(s):

Durapore Millipak Barrier 200, 0.22 m, Hydrophobic/Hydrophilic, HB/HB, Hydrophilic/Hydrophobic Durapore® 0.22 µm, Millipak® Barrier 200 Capsule

Select a Size

Change View

About This Item

eCl@ss:
32031690
UNSPSC Code:
23151806
Sign in to see your pricing
Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist


Product Name

Millipak® Barrier Capsule Media: Durapore® 0.22 µm, Chemistry: Hydrophobic/Hydrophilic Polyvinylidene Fluoride (PVDF) Inlet: 9/16 in. Hose Barb Outlet: 9/16 in. Hose Barb

Quality Segment

material

PVDF membrane, polycarbonate housing

description

hydrophobic/hydrophilic

sterility

non-sterile

sterilization compatibility

autoclavable compatible
gamma compatible

product line

Millipak®

feature

hydrophilic, hydrophobic

manufacturer/tradename

Durapore®

parameter

0.7 bar max. differential pressure (10 psid) at 25 °C (Reverse), 1.7 bar max. differential pressure (25 psid) at 80 °C (Forward), 4.1 bar max. differential pressure (60 psid) at 25 °C (Forward)

technique(s)

sterile filtration: suitable

L

6.1 in.

W

3.0 in.

filtration area

1000.0 cm2

inlet connection diam.

9/16 in.

outlet connection diam.

9/16 in.

impurities

<0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

matrix

Durapore®

pore size

0.22 μm

bubble point

≥1280 mbar (18.5 psi), nitrogen with 70/30% IPA/water at 23 °C

fitting

inlet hose barb, outlet hose barb, (14 mm (9/16 in.) Hose Barb Inlet and Outlet)

General description

Device Configuration: Capsule

Preparation Note

This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Analysis Note

Bacterial Retention
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.

Other Notes

Directions for Use

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.

Legal Information

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
MILLIPAK is a registered trademark of Merck KGaA, Darmstadt, Germany


Still not finding the right product?




Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Don't see the Right Version?

If you require a particular version, you can look up a specific certificate by the Lot or Batch number.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library