Select a Size
About This Item
AC/DC input
100 - 230 V, 50 - 60 Hz
Quality Level
product line
AFS®
packaging
pkg of 1 unit
parameter
15 L/hr make-up flow rate, 2 L/min distribution flow rate, 2 L/min product water instant delivery rate, 28 kg operating weight (61.7 lb), 300 L/day max. usage
resistivity
>15 MΩ-cm, at 25°C
system H × W × depth
58.5 cm (23 in.) × 26.8 cm (10.6 in.) × 42.6 cm (16.8 in.)
total organic carbon (TOC) residue
<30 ppb
input
potable tap water
output
clinical laboratory reagent water (CLRW) (that complies with the CLSI guideline)
General description
Application
Features and Benefits
- State-of-the-art Elix® electrodeionization (EDI) technology is combined with other complementary water purification techniques to provide consistent water quality meeting CLRW standards
- Elix® EDI technology uses permanently self-regenerating ion exchange resins to ensure predictable running costs
- Fulfills accreditation needs (e.g. CAP 15189SM accreditation to the ISO 15189:2022 standard) with full monitoring and automatic water quality archiving capabilities. Up to 6 months of information can be stored for reliable traceability
- Patented Elix® technology provides constant quality water without the need for softeners or conditioning cartridges, which means reduced maintenance ― and less analyzer downtime
- Milli-Q® Service Plans offer a range of support, including preventive maintenance visits and online contract management
- A small footprint allows installation wherever it’s convenient: on the wall, on or under the bench, or on a cart
Other Notes
- The indicated price currently corresponds to the above catalog number only; it is subject to change at any time without notice.
- This catalog number does not include consumables, accessories and other products and/or services that may be required for the installation and the correct operation of the water purification unit over time.
- Because our customers each have specific purified water needs, we recommend that you have your purchase validated by one of our specialists. We remain at your disposal to advise you, do not hesitate to contact us!
- Organism Retention: Microorganisms
- Mode of Action: Filtering and UV
- Application: Clinical laboratories
- Intended Use: Water purification
- Instructions for Use: This item provides water filtered through 0.22 ?m sterilizing-grade membranes and exposed to bactericidal UV lamp. Please refer to system equipment user guide section "System" (provided on USB key in the North American configuration version).
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state, and local regulations.
Legal Information
Disclaimer

Still not finding the right product?
Explore all of our products under AFS® Water Purification System
Regulatory Information
This item has
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Articles
Explore how high-quality water is essential for accurate clinical chemistry assays and complying with norms and guidelines (e.g. CLSI).
Critical points for laboratories to consider when assessing bacteria levels in purified water and attempting to optimize the microbiology process.
Global Trade Item Number
| SKU | GTIN |
|---|---|
| ZAFS15EWW | 04053252688744 |