Merck
CN
HomePhotometry & ReflectometryAnalytical Method: Analytical Quality Assurance, Standard for Free Chlorine (DIN)

Analytical Method: Analytical Quality Assurance, Standard for Free Chlorine (DIN)

Reagents:
Potassium iodate volumetric standard (Product No. 102404)
Potassium iodide (Product No. 105043) for analysis
Sulfuric acid (Product No. 109072 ) 0.5 mol/L
Sodium hydroxide solution (Product No. 109136) 2 mol/L
Water (Product No. 116754) for analysis

Preparation of a KlO3 stock solution:
Dissolve 1.006 g of KlO3 in 250 mL of water for analysis in a calibrated or conformity-checked 1,000-mL volumetric flask. Subsequently make up to the mark with water for analysis.

Preparation of a KlO3/Kl standard solution:
Transfer 15.0 mL of the KlO3 stock solution to a calibrated or conformity-checked 1,000-mL volumetric flask, add approx. 1 g of Kl and make up to the mark with water for analysis. 1 mL of this solution is equivalent to 0.015 mg of free chlorine.

Preparation of the chlorine standard solution:
Pipette 20.0 mL KlO3/Kl standard solution into a calibrated or conformity-checked 100-mL volumetric flask, add
2 mL of H2SO4 0.5 mol/L, leave to stand for 1 min, and then add NaOH 2 mol/L dropwise (approx. 1 ml) until the solution just loses its color. Subsequently make up the solution to the mark with water for analysis. The concentration of the solution is 3.00 mg/L free chlorine.

Stability:
The KlO3 stock solution remains stable for 4 weeks when stored in a cool place. The KlO3/Kl standard solution can be used for 5 hours when stored in a cool place. The dilute iodine standard solution is not stable and must be used immediately.

Materials

Loading
Sign In To Continue

To continue reading please sign in or create an account.

Don't Have An Account?