Photometric Determination of Total Nitrogen in Soil Following Extraction with Calcium Chloride

Introduction
Plants require nitrogen as a nutrient for their healthy growth and development, and nitrogen is imperative for protein synthesis. Three major sources of nitrogen for plants are ammonia (NH3), nitrates (NO3-), and nitrites (NO2-). Extensively, ammonia and nitrates present in fertilizer and manure-amended soil function as major nitrogen suppliers for plants. One parameter of soil health for growing crops is the concentration of total nitrogen and hence it is important to test total nitrogen in soils. The Kjeldahl method, or Kjeldahl digestion is a widely used analytical method to determine organic and ammonium nitrogen1. In this application note, we detail an alternative, quantitative photometric determination of total nitrogen test of soils (not just organic and ammonium nitrogen) with either dimethyl phenol (DMP) or a benzoic acid derivative, depending on the used test kit, after extraction with calcium chloride solution and digestion by Koroleff’s method.
Experimental
This application note details the photometric determination of total Nitrogen in soils after extraction and sample digestion.
Method
Organic and inorganic nitrogen compounds are transformed into nitrate according to Koroleff’s method by treatment with an oxidizing agent in a thermoreactor.
For test 1.14537: In concentrated sulfuric acid, this nitrate reacts with a benzoic acid derivative to form a red nitro compound that is determined photometrically.
The digestion is analogous to DIN EN ISO 11905-1.
For tests 1.14763 and 1.00613: In a solution acidified with sulfuric and phosphoric acid, this nitrate reacts with 2,6-dimethylphenol (DMP) to form 4-nitro-2,6-dimethylphenol that is determined photometrically.
The method (digestion and determination) corresponds to ISO 23697-1.
The digestion is analogous to EN ISO 11905-1.
The determination of nitrate is analogous to DIN 38405-9. photometrically.
Measuring Range
Applicable Sample
Soil samples
Reagents, Instruments and Materials
Test /Reagents Kit(s)
For the measurement, one of the following Spectroquant® test kits is necessary:
- Spectroquant® Nitrogen Cell Test (1.14537)
- Spectroquant® Nitrogen Cell Test (1.14763)
- Spectroquant® Nitrogen Cell Test (1.00613)
Instrument(s) & Devices
For the measurement, one of the following Spectroquant® photometers is necessary.
- Spectroquant® VIS Spectrophotometer Prove 100 Plus (1.73026)
- Spectroquant® UV/VIS Spectrophotometer Prove 300 Plus (1.73027)
- Spectroquant® UV/VIS Spectrophotometer Prove 600 Plus (1.73028)
- Spectroquant® Colorimeter Move 100 (1.73632)
Note: Also, legacy Spectroquant® instruments are suitable.
For the digestion, one of the following devices is necessary.
- Spectroquant® Thermoreactor TR 320 (1.71200)
- Spectroquant® Thermoreactor TR 420 (1.71201)
- Spectroquant® Thermoreactor TR 620 (1.71202)
Software for Data transfer
- Optional Spectroquant® Prove Connect to LIMS software package (Y.11086) to transfer your data into an existing LIMS system.
Instrument Accessories
- Rectangular cells 10 mm (1.14946)
- Rectangular cells 20 mm (1.14947)
- Rectangular cells 50 mm (1.14944)
Other Reagents and Accessories
- Calcium chloride dihydrate for analysis (1.02382)
- Water for analysis (1.16754)
- Analytical balance
- Standard laboratory glassware (e.g., glass beakers) and pipettes
- Shaker or stirring plate
- Spatula
- Drying kiln
- Charcoal activated for soil test
- Folded filter
Analytical Procedure
Reagent Preparation
- Calcium chloride solution 0.025 mol/L: Dissolve 3.68 g of calcium chloride dihydrate for analysis with 1 L of water to prepare a 0.025 mol/L calcium chloride solution.
Sample Preparation
- In a glass bottle mix 50 g of naturally moist sample, free from coarse stones, with 100 mL of a calcium chloride solution 0.025 mol/L.
- Add a spatula-tip full of charcoal activated for soil tests and shake the closed bottle in a shaking machine for 1 hour (alternative stir in a beaker). Let the suspension settle and filter it through a folded filter.
- For determination of the water-content dry a similar sample to constant weight in the drying kiln at 105 °C.
Using Cat. No. 1.14537: Procedure and Measurement
For more information on the measurement, see the packaging insert for the test.
Procedure
- Pipette 10 mL Pretreated sample into an empty cell.
- Add 1 level blue microspoon (in the cap of the N-1K bottle) Reagent N-1K and mix.
- Add 6 drops Reagent N-2K, close cell tightly and mix. Hold the bottle vertically while adding the reagent.
- Heat the cell at 120 °C in the preheated thermoreactor for 1 hour.
- Allow the closed cell to cool to room temperature in a test-tube rack. Do not cool with cold water!
- Shake the cell briefly after 10 min. (Turbidity or precipitation frequently occurs in the digestion solution.)
- Place 1 level microspoon (in the cap of the N-3K bottle) Reagent N-3K into a reaction cell, immediately close the cell tightly, and shake vigorously for 1 min.
- Very slowly and carefully allow 1.5 mL digested, cooled sample (use clear supernatant or filtrate in the event of turbidity or precipitation) to run from the pipette down the inside of the tilted reaction cell onto the reagent (Wear eye protection! The cell becomes hot!). Immediately close the cell tightly and mix briefly. The cell must be held only by the screw cap!
- Leave the hot cell to stand for 10 min (reaction time). Do not cool with cold water!
- Measure the sample in the photometer.
Measurement
- For photometric measurement the cells must be clean. Wipe, if necessary, with a clean dry cloth.
- Measurement of turbid solutions yields false-high readings.
- The color of the measurement solution remains stable for at least 60 min after the end of the reaction time stated above.
Hints for Measurement
- It is recommended to zero the method each new working day. To do this, open the method, either by manually selecting the method or by inserting a barcoded cell. Tap the <Settings> button and select the <ZERO ADJUSTMENT> menu item. After prompting, insert the 16 mm zero cell through the corresponding opening. The zero adjustment is performed automatically. Confirm the performance of the zero-adjustment procedure by clicking on <OK>.
- After the zero has been performed, insert the barcoded Spectroquant® round cell through the corresponding opening, ensuring that the white position mark on the cell is aligned with the positioning mark on the spectrophotometer. The measurement starts automatically.
- Read off the result in mg/L from the display.
Hint: The above written measurement description is only valid for the Spectroquant® Prove (plus) series photometer. If a different instrument is used, please consult the corresponding instrument manual for more details on how to perform the measurement.
Using Cat. No. 1.00613: Procedure and Measurement
For more information on the measurement, see the packaging insert for the test.
Procedure
- Pipette 10 mL Pretreated sample into an empty cell.
- Add 1 level blue microspoon (in the cap of the N-1K bottle) Reagent N-1K and mix.
- Add 6 drops Reagent N-2K, close cell tightly and mix. Hold the bottle vertically while adding the reagent.
- Heat the cell at 120 °C in the preheated thermoreactor for 1 hour.
- Allow the closed cell to cool to room temperature in a test-tube rack. Do not cool with cold water!
- Shake the cell briefly after 10 min. (Turbidity or precipitation frequently occurs in the digestion solution.)
- Pipette 1.0 mL digested, cooled sample (use clear supernatant or filtrate in the event of turbidity or precipitation) into a reaction cell. Do not mix content!
- Add 1.0 mL Reagent N-3K with a pipette, close the cell tightly and mix (Wear eye protection! The cell becomes hot! The cell must be held only by the screw cap!
- Leave the hot cell to stand for 10 min (reaction time). Do not cool with cold water!
- Measure the sample in the photometer.
Measurement
- For photometric measurement the cells must be clean. Wipe, if necessary, with a clean dry cloth.
- Measurement of turbid solutions yields false-high readings.
- The color of the measurement solution remains stable for 30 min after the end of the reaction time stated above. (After 60 mins the measurement value would have increased by 5%.)
Hints for Measurement
- It is recommended to zero the method for each new working day. To do this, open the method by inserting the barcode, tap the <Settings> button and select the <ZERO ADJUSTMENT> menu item. Fill the same cell which will be used for the sample measurement with distilled water. After prompting, insert the filled rectangular cell into the cell compartment. The zero adjustment is performed automatically. Confirm the performance of the zero-adjustment procedure by clicking on <OK>.
- After the zero adjustment, fill the measurement sample into the same or a matched rectangular cell and insert the cell into the cell compartment. The measurement starts automatically.
- Read off the result in mg/L from the display.
Hint: The above written measurement description is only valid for the Spectroquant® Prove (plus) series photometer. If a different instrument is used, please consult the corresponding instrument manual for more details on how to perform the measurement.
Using Cat. No. 1.14763: Procedure and Measurement
For more information on the measurement, see the packaging insert for the test.
Procedure
- Pipette 1 mL Pretreated sample into an empty cell.
- Add 9 mL water (recommended is Cat. No. 1.16745, or use distilled water).
- Add 1 level blue microspoon (in the cap of the N-1K bottle) Reagent N-1K and mix.
- Add 6 drops Reagent N-2K, close cell tightly and mix. Hold the bottle vertically while adding the reagent.
- Heat the cell at 120 °C in the preheated thermoreactor for 1 hour.
- Allow the closed cell to cool to room temperature in a test-tube rack. Do not cool with cold water!
- Shake the cell briefly after 10 min. (Turbidity or precipitation frequently occurs in the digestion solution.)
- Pipette 1.0 mL digested, cooled sample (use clear supernatant or filtrate in the event of turbidity or precipitation) into a reaction cell. Do not mix content!
- Add 1.0 mL Reagent N-3K with a pipette, close the cell tightly and mix (Wear eye protection! The cell becomes hot! The cell must be held only by the screw cap!
- Leave the hot cell to stand for 10 min (reaction time). Do not cool with cold water!
- Measure the sample in the photometer.
Measurement
- For photometric measurement the cells must be clean. Wipe, if necessary, with a clean dry cloth.
- Measurement of turbid solutions yields false-high readings.
- The color of the measurement solution remains stable for 30 min after the end of the reaction time stated above. (After 60 mins the measurement value would have increased by 5%.)
Hints for Measurement
- It is recommended to zero the method for each new working day. To do this, open the method by inserting the barcode, tap the <Settings> button and select the <ZERO ADJUSTMENT> menu item. Fill the same cell which will be used for the sample measurement with distilled water. After prompting, insert the filled rectangular cell into the cell compartment. The zero adjustment is performed automatically. Confirm the performance of the zero-adjustment procedure by clicking on <OK>.
- After the zero adjustment, fill the measurement sample into the same or a matched rectangular cell and insert the cell into the cell compartment. The measurement starts automatically.
- Read off the result in mg/L from the display.
Hint: The above written measurement description is only valid for the Spectroquant® Prove (plus) series photometer. If a different instrument is used, please consult the corresponding instrument manual for more details on how to perform the measurement.
Analytical Quality Assurance
Analytical quality assurance (AQA) is recommended before each measurement series.
To check the photometric measurement system (test reagents, measurement device, handling) and the mode of working, the nitrogen (total) standard solutions (see section 5 of the respective packaging insert) or Spectroquant® CombiCheck 50 (Cat. No. 1.14695) or CombiCheck 70 (Cat. No. 1.14689) respectively can be used. Besides a standard solution with 5.0 mg/L N or 50 mg/L N, respectively, CombiChecks also contains an addition solution for determining sample-dependent interferences (matrix effects).
Sample-dependent interferences (matrix effects) can be determined by means of standard addition or dilution.
To view additional notes, visit SigmaAldrich.com/qa-test-kits.
Calculation
Total nitrogen content in mg/kg N = analysis value in mg/L N x 2
Related Products
Test Kits & Strips
Instruments, Software & Accessories
Reagents
Reference Materials
References
To continue reading please sign in or create an account.
Don't Have An Account?