GDS OTV Insert

BioControl, For use with GDS, Performance verification of the GDS Rotor-Gene

登录查看公司和协议定价

选择尺寸


关于此项目

UNSPSC代码:
41116135
技术服务
需要帮助?我们经验丰富的科学家团队随时乐意为您服务。
让我们为您提供帮助
技术服务
需要帮助?我们经验丰富的科学家团队随时乐意为您服务。
让我们为您提供帮助

质量水平

保质期

1‑5 mo.

制造商/商品名称

BioControl Systems 73021

储存条件

protect from light at room temperature

技术

PCR: suitable

应用

food and beverages
pathogen testing

相容性

for use with GDS Rotor-Gene

储存温度

room temp

一般描述

The GDS OTV (Optical Temperature Verification) Insert provides automated performance verification and dynamic calibration of the GDS Rotor-Gene. The OTV system uses the chemical properties of thermochromic liquid crystals (TLC) as an absolute temperature reference. When heated, a TLC changes from opaque to transparent at a very precise temperature. The Rotor-Gene can measure this precise temperature transition point and, by comparing the measured in-tube thermal transition point to the known calibrated value, the OTV software verifies that the Rotor-Gene is within specification. If necessary, the OTV software allows the user to automatically recalibrate the instrument with a simple push of a button. A printable post-run report documents each test and any necessary recalibration.

应用

Optical Temperature Verification

其他说明

  • OTV Insert – good for 30 verification runs
  • Optical Filters for Rotor-Gene 3000 and Rotor-Gene Q, one each