产品名称
Aervent® 0.2 µm小柱, Chemistry: 0.2 µm Hydrophobic Polytetrafluoroethylene (PTFE) Configuration: 10 in. Code 7
Quality Segment
material
polypropylene support, PTFE membrane, polypropylene housing, silicone O-ring
sterility
non-sterile
sterilization compatibility
steam-in-place compatible
product line
Aervent®
feature
hydrophobic
manufacturer/tradename
Aervent®
parameter
4.1 bar max. differential pressure (60 psid) at 23 °C (Reverse), 5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)
technique(s)
gas filtration: suitable
W
2.7 in.
cartridge nominal length
10 in. (25 cm)
device L
10.0 in. (25.0 cm)
diam.
6.9 cm (2.7 in.)
filtration area
0.65 m2
impurities
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
gravimetric extractables
≤35 mg/cartridge
matrix
Aervent®
pore size
0.2 μm pore size
input
sample type gas
bubble point
≥1100 mbar (16 psig), nitrogen with 70/30% IPA/water at 23 °C
cartridge code
Code 7
General description
设备配置:纯化柱
Features and Benefits
每个滤芯在85 Nm3/h (50 SCFM)流速下,可截留108–1010空斑形成单位的ΦX-174病毒颗粒气溶胶。
Packaging
双层易开袋
Preparation Note
本产品采用符合21 CFR 210.3(b)(6)中规定的“无纤维释放”过滤器标准的PTFE膜生产。
灭菌方法
30 min、145 °C条件下150次(100 次正向/50次反向)在线灭菌
30 min、145 °C条件下150次(100 次正向/50次反向)在线灭菌
Analysis Note
HydroCorr:≤2.6 bar(38 psig)压力下,10英寸滤芯的水流量≤0.75 mL/min
可提取物质量:受控室温下,在70/30% 异丙醇/水混合液中浸泡24小时后测定。
截留缺陷短波单胞菌(原名:缺陷假单胞菌)液体气溶胶21天,挑战最低值10⁷ CFU/cm²
细菌保留
使用ASTM® F838方法,对样品进行最小保留的缺陷短波单孢菌挑战浓度为 1 × 10⁷ CFU/cm² 的定量保留。
使用ASTM® F838方法,对样品进行最小保留的缺陷短波单孢菌挑战浓度为 1 × 10⁷ CFU/cm² 的定量保留。
Other Notes
使用说明
- 生物截留:微生物
- 作用机制:过滤(体积排阻)
- 应用:生物处理
- 预期用途:减少或去除微生物/生物负荷
- 使用说明:参见蒸汽灭菌&《完整性测试程序指南》和《目视检查指南》第一页
- 储存声明:干燥储存
- 处置声明:根据国家、省市和当地适用法规处置。
Legal Information
AERVENT is a registered trademark of Merck KGaA, Darmstadt, Germany
ASTM is a registered trademark of American Society for Testing and Materials
Still not finding the right product?
Explore all of our products under Aervent® 0.2 µm小柱