等级
analytical standard
质量水平
质量
traceable to primary standards (LGC)
保质期
limited shelf life, expiry date on the label
mp
334 °C (lit.)
应用
food and beverages
pharmaceutical
包装形式
neat
SMILES字符串
[K+].[O-][N+]([O-])=O
InChI
1S/K.NO3/c;2-1(3)4/q+1;-1
InChI key
FGIUAXJPYTZDNR-UHFFFAOYSA-N
正在寻找类似产品? 访问 产品对比指南
一般描述
Mettler-Toledo calibration substance ME 51143095, potassium nitrate is provided with the melting point specific to Mettler-Toledo instrument, documented under both thermodynamic mode and pharmacopeia modes.
The melting point value equals a mean of 6 to 12 measurements with a Mettler-Toledo MP90 Excellence instrument that is calibrated with primary standards. The mp value is found using Capillary method, in accordance with European Pharmacopeia (2.2.14.)
The melting point value equals a mean of 6 to 12 measurements with a Mettler-Toledo MP90 Excellence instrument that is calibrated with primary standards. The mp value is found using Capillary method, in accordance with European Pharmacopeia (2.2.14.)
应用
Potassium nitrate, Mettler-Toledo calibration substance ME 51143095 is used as a melting point calibration standard to assess the performance of Mettler-Toledo melting point instruments.
特点和优势
- Melting point traceable to a primary standard (LGC, London)
- Physically correct and heating rate dependent melting point evaluations using thermodynamic and pharmacopeia modes of testing, respectively
- Certificates of analysis and safety data sheet provided
- Uncertainty of measurement up to ± 0.5 °C
警示用语:
Warning
危险声明
危险分类
Ox. Sol. 3
储存分类代码
5.1B - Oxidizing hazardous materials
WGK
WGK 1
闪点(°F)
Not applicable
闪点(°C)
Not applicable
个人防护装备
Eyeshields, Gloves, type P3 (EN 143) respirator cartridges