grade
analytical standard
Quality Level
description
redox potential: 650 mV ±5 mV, 25°C
shelf life
limited shelf life, expiry date on the label
application(s)
environmental
pharmaceutical
format
mixture
storage temp.
2-8°C
General description
This Redox standard has a potential of 650 mV ±5 mV. It refers to the potential of the platinum electrode versus the Ag/AgCl reference electrode in 3M potassium chloride.
Application
Redox standard acts as a control standard rather than a calibration standard which controls the functionality of the sensing and reference electrodes.
Features and Benefits
- Broad range of values
- Very high specifications (±5mV)
- Detailed safety data sheet (SDS) available
- Certified with proven verifiable accuracy and uncertainty of measurement
- Guaranteed stability throughout its entire shelf life
Still not finding the right product?
Explore all of our products under 氧化还原标准品
signalword
Danger
hcodes
Hazard Classifications
Eye Dam. 1 - Met. Corr. 1 - Skin Irrit. 2
flash_point_f
Not applicable
flash_point_c
Not applicable
存储类别
12 - Non Combustible Liquids
全球贸易项目编号
| 货号 | GTIN |
|---|---|
| 87783-500ML | 04061838095435 |