Product Name
Millipore Express® SHR, Cartridge, cartridge nominal length 10 in. (25 cm), filter diam. 6.9 cm (2.7 in.), Code 7
parameter
6.9 bar max. differential pressure (100 psi) at 25 °C (Forward)
0.34 bar max. differential pressure (5 psi) at 135 °C
1 bar max. differential pressure (15 psi) at 135 °C
1.7 bar max. differential pressure (25 psi) at 80 °C
2.1 bar max. differential pressure (30 psi) at 25 °C (Reverse)
69 mbar max. differential pressure (1 psi) at 135 °C
Quality Level
material
fluoroelastomer O-ring
polyethersulfone membrane
polypropylene housing
polypropylene support
seal (Fluorocarbon rubber)
reg. compliance
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
sterility
non-sterile
sterilization compatibility
autoclavable compatible
steam-in-place compatible
product line
Millipore Express®
feature
hydrophilic
manufacturer/tradename
Millipore Express®
technique(s)
sterile filtration: suitable
L
25.4 cm (10 in.)
W
2.7 in.
cartridge nominal length
10 in. (25 cm)
filter diam.
6.9 cm (2.7 in.)
filter filtration area
0.60 m2 (6.5 ft2)
filter size
10 in.
impurities
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
matrix
Millipore Express® SHR
pore size
0.1 μm pore size
bubble point
≥2586 mbar (37.5 psig), nitrogen with 70/30% IPA/water at 23 °C
cartridge code
Code 7
Looking for similar products? Visit Product Comparison Guide
Analysis Note
TOC/Conductivity: After sterilization and a controlled water flush of 10 L, samples exhibited <500 ppb TOC per USP 643 and <1.3 μS/cm per USP 645 at 25 °C.
Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2 using ASTM® F838 methodology.
General description
Other Notes
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer wetting instructions of Filters with MILLIPORE EXPRESS® Membrane user guide
- Storage Statement: Please refer Filters with MILLIPORE EXPRESS® Membrane user guide
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations
Packaging
Preparation Note
This product was manufactured with a MILLIPORE EXPRESS® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
Legal Information
Regulatory Information
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service