Select a Size
About This Item
Product Name
MILLIPORE EXPRESS® SHR Cartridge, Chemistry: 0.1 µm Hydrophilic Polyethersulfone (PES) Configuration: 30 in. Code 7
Quality Level
material
polypropylene support, silicone seal, polyethersulfone membrane, polypropylene housing, silicone O-ring
reg. compliance
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
sterility
non-sterile
sterilization compatibility
autoclavable compatible
steam-in-place compatible
product line
Millipore Express®
feature
hydrophilic
manufacturer/tradename
Millipore Express®
parameter
0.34 bar max. differential pressure (5 psi) at 135 °C (Forward), 1.0 bar max. differential pressure (15 psi) at 135 °C (Forward), 1.7 bar max. differential pressure (25 psi) at 80 °C (Forward), 2.1 bar max. differential pressure (30 psi) at 25 °C (Reverse), 6.9 bar max. differential pressure (100 psi) at 25 °C (Forward), 69 mbar max. differential pressure (1 psi) at 135 °C (Reverse)
technique(s)
sterile filtration: suitable
L
76.2 cm (30 in.)
W
2.7 in.
cartridge nominal length
30 in. (75 cm)
diam.
20.7 cm (8.1 in.)
filtration area
1.80 m2 (19.5 ft2)
impurities
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
matrix
Millipore Express® SHR
pore size
0.1 μm pore size
input
sample type liquid
bubble point
≥2586 mbar (37.5 psig), nitrogen with 70/30% IPA/water at 23 °C
cartridge code
Code 7
General description
Packaging
Preparation Note
25 forward or 22 forward and 3 reverse SIP cycles of 30 min @ 135 °C; 25 autoclave cycles of 60 min @ 126 °C
Analysis Note
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
After sterilization and a controlled water flush of 30 L, samples exhibited less than 500 ppb TOC per USP <64> and less than 1.3 µS/cm per USP <64> at 25 °C.
Other Notes
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer wetting instructions of Filters with MILLIPORE EXPRESS® Membrane user guide
- Storage Statement: Please refer Filters with MILLIPORE EXPRESS® Membrane user guide
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Legal Information
Still not finding the right product?
Explore all of our products under MILLIPORE EXPRESS® SHR Cartridge
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Global Trade Item Number
| SKU | GTIN |
|---|---|
| CVEP73TS3 | 04053252303555 |