Select a Size
About This Item
Product Name
Opticap® XL 4 CapsuleMedia: Durapore® 0.1 µm, Chemistry: Hydrophilic Polyvinylidene Fluoride (PVDF)Inlet: 9/16 in. Hose BarbOutlet: 9/16 in. Hose BarbMax Operating Pressure: 80 psi @ 25 °C
Quality Level
material
PVDF , polypropylene , polypropylene housing, polypropylene support, polypropylene vent cap, silicone seal
reg. compliance
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
sterility
non-sterile
sterilization compatibility
autoclavable compatible
product line
Opticap® XL 4
feature
hydrophilic
manufacturer/tradename
Opticap®
parameter
≤7.5 mL/min air diffusion at 3.9 bar (56 psig) and 23 °C (in water), 1.0 bar max. differential pressure (15 psid) at 80 °C (Forward), 1.0 bar max. inlet pressure (15 psi) at 80 °C, 2.8 bar max. inlet pressure (40 psi) at 60 °C, 25 °C max. inlet temp., 3.5 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent), 5.5 bar max. differential pressure (80 psid) at 25 °C (Forward), 5.5 bar max. inlet pressure (80 psi) at 23 °C, 80 psig max. inlet pressure
technique(s)
sterile filtration: suitable
L
19.3 cm (7.6 in.)
cartridge nominal length
4 in. (10 cm)
diam.
14.5 cm (5.7 in.)
filtration area
0.19 m2
inlet connection diam.
9/16 in.
inlet to outlet W
19.3 cm (7.6 in.)
outlet connection diam.
9/16 in.
impurities
≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction), <0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
gravimetric extractables
≤10 mg/capsule
matrix
Durapore®
pore size
0.1 μm
input
sample type liquid
bubble point
≥4830 mbar (70 psig), air with water at 23 °C
fitting
1/4 in. drain/vent hose barb (with double O-ring Seal), (14 mm (9/16 in.) Hose Barb Inlet and Outlet)
General description
Packaging
Preparation Note
3 autoclave cycles of 60 min @ 126 °C; not in-line steam sterilizable
Analysis Note
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
Other Notes
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Legal Information
Disclaimer
Still not finding the right product?
Explore all of our products under Opticap® XL 4 CapsuleMedia: Durapore® 0.1 µm
Stay compliant with GMP using the Emprove® Program
Accelerate qualification, maintain compliance, and save time with Emprove® Dossiers.
Note: you will be brought to EmproveSuite.com. Your username and password are separate from SigmaAldrich.com and may differ.
Register for free and access Emprove® Dossiers within minutes. Upgrade your subscription plan to unlock advanced and customized content in the Emprove® Suite
Not finding what you are looking for? Visit the Emprove® Program Help page.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Global Trade Item Number
| SKU | GTIN |
|---|---|
| KVVLA04HH3 | 04053252355233 |