  • Validation of a HPLC method for the determination of p-nitrophenol hydroxylase activity in rat hepatic microsomes.

Validation of a HPLC method for the determination of p-nitrophenol hydroxylase activity in rat hepatic microsomes.

Journal of chromatography. B, Analytical technologies in the biomedical and life sciences (2006-02-25)
Fawzy Elbarbry, Kyle Wilby, Jane Alcorn
ABSTRACT

We report a HPLC-UV method for determination of p-nitrophenol (PNP) hydroxylation to 4-nitrocatechol (4NC) as a marker for CYP2E1 activity in rat hepatic microsomes. Proteins were precipitated by addition of 50 microL phosphoric acid (50%, v/v in water) to 500 microL microsomal suspensions. Following vortex mixing and centrifugation the supernatant (20 microL) was injected onto a Supelcosil C(18) column (150 mm x 4.6 mm, 5 microm), and mobile phase (22% acetonitrile, 0.1% trifluoroacetic acetic acid, 0.5% triethylamine) delivered at 1.0 mL/min produced resolved peaks for internal standard, 4NC, and PNP in < 11 min. Calibration curves were linear (r(2) = 0.999) from 0.1 to 40 microM with intra- and inter-day precision < 12% and accuracy >90%. The method's improved sensitivity (LOQ = 0.1 microM) and minimal sample processing allowed rapid monitoring of PNP hydroxylase activity in fetal, neonatal, juvenile, and adult rat livers.

MATERIALS
Product Number
Brand
Product Description

Supelco
SUPELCOSIL LC-NH2-NP HPLC Column, 5 μm particle size, L × I.D. 25 cm × 4.6 mm
Supelco
SUPELCOSIL LC-NH2-NP Supelguard Cartridge, 5 μm particle size, L × I.D. 2 cm × 4 mm