产品名称
Aervent® 0.2 µm小柱, Chemistry: 0.2 µm Hydrophobic Polytetrafluoroethylene (PTFE) Configuration: 30 in. Code 0
Quality Segment
material
polypropylene support, PTFE membrane, polypropylene housing, silicone O-ring
sterility
non-sterile
sterilization compatibility
steam-in-place compatible
product line
Aervent®
feature
hydrophobic
manufacturer/tradename
Aervent®
parameter
4.1 bar max. differential pressure (60 psid) at 23 °C (Reverse), 5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)
technique(s)
gas filtration: suitable
W
2.7 in.
cartridge nominal length
30 in. (76 cm)
device L
30.0 in. (75.0 cm)
diam.
6.9 cm (2.7 in.)
filtration area
1.95 m2
impurities
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
gravimetric extractables
≤105 mg/cartridge
matrix
Aervent®
pore size
0.2 μm pore size
input
sample type gas
bubble point
≥1100 mbar (16 psig), nitrogen with 70/30% IPA/water at 23 °C
cartridge code
Code 0 (2-222; O-rings)
General description
Device Configuration: Cartridge
Features and Benefits
Rated for aerosol retention of a total challenge of 108–1010 plaque forming units of ΦX-174 virus particles at 85 Nm3/h (50 SCFM) per cartridge.
Packaging
Double Easy-Open bag
Preparation Note
Sterilization Method
150 (100 forward/50 reverse) SIP cycles of 30 min @ 145 °C
150 (100 forward/50 reverse) SIP cycles of 30 min @ 145 °C
This product was manufactured with a PTFE membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
Analysis Note
Bacterial Retention
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
HydroCorr: ≤2.3 mL/min @ 2.6 bar (38 psig) per 30 in cartridge
Gravimetric Extractables: after 24 hours in a 70/30% IPA/water mixture at controlled room temperature
Retains Brevundimonas diminuta liquid aerosol for 21 days at a minimum challenge of 10⁷ CFU/cm²
Other Notes
Directions for Use
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer Steam Sterilization & Integrity Testing Procedures guide and first page of Visual Inspection Guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Legal Information
AERVENT is a registered trademark of Merck KGaA, Darmstadt, Germany
ASTM is a registered trademark of American Society for Testing and Materials
Still not finding the right product?
Explore all of our products under Aervent® 0.2 µm小柱