material
molded polyetherimide ((21 CFR part 177.1595)), platinum cured silicone seal ((21 CFR part 177.2600))
sterility
sterile; γ-irradiated (⩾25kGy)
product line
Lynx®
parameter
0.7 bar max. pressure (10 psi) ((during actuation)), 23 L/min flow rate at 0.6 bar (9 psi), 38 L/min flow rate at 1.8 bar (26 psi), 4-80 °C operating temperature ( (39-176 °F) (SIP : 4X at 145°C/293°F for 30 min. Closed), 6.8 bar max. pressure (100 psi) ((operating pressure)), 8 L/min flow rate at 0.1 bar (1.5 psi)
size
95.3 -84.1 mm (3.8 -3.3 in.) × 95.3 -91.4 mm (3.8-3.6 in.)
fitting
1.5 in. Tri-Clover outlet, 1/2 in. outlet hose barb
application(s)
pharma/biopharma processes
正在寻找类似产品? 访问 产品对比指南
General description
The Lynx® ST (Steam-To) connector is a reliable and validated sterile connector enabling a sterile transfer of process fluids. One of many Mobius® single-use bioprocessing solutions, it is designed to connect steamable hard piped processing systems to sterilized disposable flow paths. The Lynx® ST connector is available in a wide range of sizes and configurations, from sampling to intermediate additions, to bulk liquid transfer into and out of the stainless steel tanks or vessels.
Application
Sterile transfer of process fluids
Packaging
10 units / pack
Preparation Note
Steam @ 145 °C for 30 min (closed) 4 times Autoclave: 60min at 126°C
Legal Information