产品线
SeQuant®
质量水平
参数
temp. range at ambient-70 °C
技术
HPLC: suitable
LC/MS: suitable
长度 × 内径
50 mm × 0.3 mm
基质
fully porous particle particle platform
基质活性基团
ZIC-cHILIC bonding phase
粒径
3 μm
孔径
100 Å pore size
工作pH范围
2-8
分离技术
hydrophilic interaction (HILIC)
正在寻找类似产品? 访问 产品对比指南
一般描述
The SeQuant® ZIC®-cHILIC capillary columns have a zwitterionic stationary phase covalently attached to porous silica. The permanent and hydrophilic zwitterion functionality makes the column suitable for Hydrophilic Interaction Liquid Chromatography (HILIC). The capillary and microbore columns (0.3–1.0 mm I.D.) are made from stainless steel that has been internally glass-lined and has 10–32 UNF female fittings. The frits have a porosity to retain 3–5 μm particles and are made from stainless steel.
储存及稳定性
Store columns as shipped: Acetonitrile/5 mM ammoniumacetate, pH 6.8; 80 : 20 (v/v). Dispose the column according to local authorities andregulations.
分析说明
These columns are are designed forexcellent HPLC and LC-MS separation of polar, hydrophilic compounds. SeQuant® ZIC®-cHILIC, with phosphorylcholine functional groups, provides you with complementary selectivity for easier method development for analytes that have been difficult to separate by previous types of HPLC columns operated in reversed-phase or HILIC mode.
法律信息
SeQuant is a registered trademark of Merck KGaA, Darmstadt, Germany
ZIC is a registered trademark of Merck KGaA, Darmstadt, Germany
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系客户支持