1.17883
Magnetic perforated lid Microbial Air Sampler

for use with MAS-100 Sirius® with 90 mm settle plates at 100 SLPM flow rate
material
stainless steel lid
product line
MAS-100®
feature
no. of holes 300 x 0.6 mm
packaging
pkg of 1 unit
manufacturer/tradename
(Perforated Lid Sirius Type A 90 mm)
parameter
100 L/min flow rate
application(s)
air monitoring
compatibility
for use with MAS-100® Sirius®
Looking for similar products? Visit Product Comparison Guide
General description
The perforated lid A 90 mm for MAS-100 Sirius Microbial Air Samplers is designed to fit tightly on the instrument by easy and quick magnetic closure without the need for twisting.
It is made from 316L stainless-steel which has good resistance towards a broad range of disinfectants including vaporized hydrogen peroxide and is suitable for autoclaving.
The magnetic lid type A 90 mm has a sieve geometry of 300 holes with a diameter of 0.6 mm and is suitable for use with 90 mm agar plates. This sieve configuration delivers a nominal d50 value of 1.1 µm at a flow rate of 100 SLPM (standard liter per minute).
Each lid is marked with laser-based marking without significantly altering the properties of the metal surface. The information of the unique marking includes the identification code of the perforated lid as well as the serial number in barcode and readable format.
The perforated lid A 90 mm for MAS-100 Sirius® Microbial Air Samplers is designed to fit tightly on the instrument by easy and quick magnetic closure without the need for twisting.
It is made from 316L stainless-steel which has good resistance towards a broad range of disinfectants including vaporized hydrogen peroxide and is suitable for autoclaving.
The magnetic lid type A 90 mm has a sieve geometry of 300 holes with a diameter of 0.6 mm and is suitable for use with 90 mm agar plates. This sieve configuration delivers a nominal d50 value of 1.1 µm at a flow rate of 100 SLPM (standard liter per minute).
Each lid is marked with laser-based marking without significantly altering the properties of the metal surface. The information of the unique marking includes the identification code of the perforated lid as well as the serial number in barcode and readable format.
Application
Features and Benefits
- Independently validated efficiency acc. ISO 14698 and EN 17141 at a flow rate of 100 SLPM with ICR settles plates
- Stainless steel for optimized cleaning and sterilization
- Unique marking allows manual or digital tracking
Legal Information
Regulatory Information
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Related Content
Discover the MAS-100 Sirius® Microbial Air Sampler, designed for reliable air monitoring. Explore features, training, and support services to enhance your lab's compliance and efficiency.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service