713-1329
GC Analysis of Impurities in MTBE on Petrocol® DH
suitable for GC
technique(s)
gas chromatography (GC): suitable
test parameters
column: Petrocol DH, 100 m × 0.25 mm I.D., 0.50 μm (24160-U)
oven: 50 °C (13 min), 10 °C/min to 180 °C
inj. temp.: 250 °C
detector: FID, 310 °C
carrier gas: helium, 20 cm/sec @ 35 °C
injection: 1 μL, 200:1 split
sample: MTBE Contaminants Mix A (47942)
application(s)
chemicals and industrial polymers