US EPA Method 8082: GC Analysis of Aroclor 1254 (PCB) on SLB®-5ms and Equity-1701 (15 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis
suitable for GC
technique(s)
gas chromatography (GC): suitable
test parameters
column: SLB-5ms, 15 m x 0.10 mm I.D., 0.10 μm (28466-U)
column: Equity-1701, 15 m x 0.10 mm I.D., 0.10 μm (28343-U)
oven: 80 °C (0.5 min), 30 °C/min to 280 °C (3 min)
inj. temp.: 250 °C
detector: μ-ECD, 300 °C
carrier gas: hydrogen, 40 cm/sec, constant
injection: 1 μl, splitless (0.75 min)
liner: 4 mm I.D., single taper
sample: Ar1254 (44808) and surrogates (48460) at 500 ppb and 50 ppb in n-hexane
application(s)
environmental