G003882
US EPA Method 8015 (modified): GC Analysis of Gasoline Range Organics (GRO) on Equity-1 (15 m x 0.10 mm I.D., 0.10 μm), Direct Injection, Fast GC Analysis
suitable for GC
technique(s)
gas chromatography (GC): suitable
test parameters
column: Equity-1, 15 m × 0.10 mm I.D., 0.10 μm (28039-U)
oven: 75 °C, 40 °C/min to 110 °C, 7.5 °C/min to 190 °C
inj. temp.: 200 °C
detector: FID, 250 °C
carrier gas: hydrogen, 57 cm/sec @ 75 °C
injection: 0.5 μL, 200:1 split
liner: 4 mm I.D., split, cup design
sample: UST Modified GRO Mix, each analyte at 1000 ppm in methanol
application(s)
environmental