G006323

GC Analysis of PCBs in Cod Liver Oil (Spiked) on SLB®-5ms after SPE using Supelclean EZ-POP NP and Supelclean LC-Si

application for SPE, suitable for GC

Sign Into View Organizational & Contract Pricing

Select a Size


Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist
Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist

technique(s)

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

test parameters

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: cod liver oil spiked with PCBs (at 10 ng/g)
SPE tube/cartridge: Supelclean EZ-POP NP, 12 mL, Pk 20 (54341-U)
condition: 10 mL acetone, 10 min dry at 20" mercury
sample addition: 0.5 g sample, weighed directly on cartridge
elution: 15 mL acetonitrile
eluate post-treatment: back extract acetonitrile eluent with 2 x 10 mL portions of hexane, evaporate hexane extract to approximately 1 mL at 40 ºC under nitrogen
SPE tube/cartridge: Supelclean LC-Si, 100 mg/1 mL (504041)
condition: 4 mL acetone, 10 min dry at 20" mercury, then 4 mL hexane
sample addition: 1 mL of resulting extract from above
elution: 5 mL hexane
eluate post-treatment: evaporate to 1 mL at 40 ºC under nitrogen
column: SLB-5ms 20m x 0.18mm, 0.18 μm (28564-U)
oven: 75 °C (1 min), 12 °C/min to 340 °C (20 min)
inj. temp.: 250 °C
detector: μECD, 340 °C
carrier gas: hydrogen, 1.2 mL/min
injection: 1 μL, splitless (0.75 min)
liner: 4 mm I.D., split/splitless type, wool packed single taper FocusLiner design

suitability

application for SPE

application(s)

food and beverages

Analysis Note

PCBs are lipophilic and they have a tendency to bioaccumulate and, through bioaccumulation, may pose a risk to humans. When PCB-containing substances are cleaned or dumped into rivers or oceans, PCBs may find their way into aquatic lifeforms such as fish. Fish oil, which is high in levels of the beneficial fatty acids EPA and DHA, is a popular nutritional supplement. Due to the health risk of exposure to PCBs, fish oil is often analyzed for the presence of these compounds to ensure its safety. In this application, a new extraction/cleanup method was developed for PCBs from pure cod liver oil (which was spiked with PCBs) using the EZ-POP NP SPE cartridge.

Legal Information

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supelclean is a trademark of Sigma-Aldrich Co. LLC

Analyte

Description

    2,4,4′,-Trichlorobiphenyl (No. 28)
    2,2′,5,5′-Tetrachlorobiphenyl (No. 52)
    2,2′,4,5,5′-Pentachlorobiphenyl (No. 101)
    3,4,4′,5-Tetrachlorobiphenyl (No.81)
    3,3′,4,4′-Tetrachlorobiphenyl (No. 77)
    2′,3,4,4′,5-Pentachlorobiphenyl (No. 123)
    2,3′,4,4′,5-Pentachlorobiphenyl (No. 118)
    2,3,4,4′,5-Pentachlorobiphenyl (No. 114)
    2,2′,4,4′,5,5′-Hexachlorobiphenyl (No. 153)
    2,3,3′,4,4′-Pentachlorobiphenyl (No. 105)
    2,2′,3,4,4′,5′-Hexachlorobiphenyl (No. 138)
    3,3′,4,4′,5-Pentachlorobiphenyl (No. 126)
    2,3′,4,4′,5,5′-Hexachlorobiphenyl (No. 167)
    2,3,3′,4,4′,5-Hexachlorobiphenyl (No. 156)
    2,3,3′,4,4′,5′-Hexachlorobiphenyl (No. 157)
    2,2′,3,4,4′,5,5′-Heptachlorobipheny (No. 180)
    3,3′,4,4′,5,5′-Hexachlorobiphenyl (No. 169)
    2,3,3′,4,4′,5,5′-Heptachlorobiphenyl (No. 189)
    2,2′,3,3′,4,4′,5′5′,6,6′-Decachlorobiphenyl (No. 209)
    See All (19)

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documents section.

If you need assistance, please contact Customer Support

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service