LC/MS/MS Analysis of Estradiol (Underivatized) from Patient Serum on Titan C18, 10 cm Column

application for LC-MS

Sign Into View Organizational & Contract Pricing

Select a Size


Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist
Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist

technique(s)

LC/MS: suitable

test parameters

column: Titan C18, 10 cm x 2.1 mm I.D., 1.9 μm particles (577124-U)
mobile phase: [A] 0.2 mM ammonium fluoride; [B] methanol
gradient: 40 to 82%B in 4.6 min; to 100%B in 0.05 min; held for 1.15 min; to 40% B in 0.05 min; held for 1.65 min
flow rate: 0.4 mL/min
column temp.: 40 °C
detector: MS, ESI(-), MRM transitions listed on figure
injection: 20 μL
sample: patient serum sample containing estradiol, 56 pmol/L

suitability

application for LC-MS

application(s)

clinical

Analysis Note

A fast and robust LC/MS/MS method was developed for the measurement of estradiol in human serum using Titan C18 1.9 μm columns. Ultra high purity LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant certified reference material grade standards provided reliable quantification.

Legal Information


Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documents section.

If you need assistance, please contact Customer Support

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service