G006437
LC/MS/MS Analysis of Estradiol (Underivatized) from Patient Serum on Titan C18, 10 cm Column
application for LC-MS
technique(s)
LC/MS: suitable
test parameters
column: Titan C18, 10 cm x 2.1 mm I.D., 1.9 μm particles (577124-U)
mobile phase: [A] 0.2 mM ammonium fluoride; [B] methanol
gradient: 40 to 82%B in 4.6 min; to 100%B in 0.05 min; held for 1.15 min; to 40% B in 0.05 min; held for 1.65 min
flow rate: 0.4 mL/min
column temp.: 40 °C
detector: MS, ESI(-), MRM transitions listed on figure
injection: 20 μL
sample: patient serum sample containing estradiol, 56 pmol/L
suitability
application for LC-MS
application(s)
clinical
Analysis Note