LC/MS Analysis of 3-Methoxytyramine and Metanephrines in Horse Urine on Ascentis® Express Biphenyl after On-line SPE using Supel™Genie RP-Amide Cartridge
application for LC-MS
technique(s)
LC/MS: suitable
test parameters
column: Ascentis® Express Biphenyl 10 cm x 2.1 mm I.D., 2.7 μm particles (64065-U)
SPE tube/cartridge: Supel™ Genie On-line cartridge: RP-Amide 2 cm x 4.0 mm I.D. (55516-U)
mobile phase: [A] 0.1% formic acid in water; [B] 0.1% formic acid in methanol
gradient: 20 to 100% B in 10 min; held at 100% B for 1 min
flow rate: 0.5 mL/min
column temp.: 25 °C
detector: MS, ESI(+), MRM, m/z 184>166, 198>180, 168>151
injection: 5 μL
sample: 200 ng/mL in horse urine (Sample loading and washing: 0.3 mL/min for 2 min, then the valve switches to in-line with HPLC column before sample loading the trap is equilibrated with the loading solvent for 2.5 min. Sample loading solvent: 5% methanol)