  • A "de-streaking" method for two-dimensional electrophoresis using the reducing agent tris(2-carboxyethyl)-phosphine hydrochloride and alkylating agent vinylpyridine.

A "de-streaking" method for two-dimensional electrophoresis using the reducing agent tris(2-carboxyethyl)-phosphine hydrochloride and alkylating agent vinylpyridine.

Proteomics (2005-04-23)
Fengju Bai, Sheng Liu, Frank A Witzmann
ABSTRACT

Optimal isoelectric focusing in the alkaline region remains a challenge in two-dimensional gel electrophoresis (2-DE), though various attempts had been made to reduce basic end streaking. The present study reports the application of a novel reduction and alkylation step prior to 2-DE analysis using tris(2-carboxyethyl)-phosphine hydrochloride as a reducing agent and vinylpyridine as an alkylating agent. This simple sample preparation approach effectively eliminates basic end streaks, thereby enabling the analysis and identification of more protein spots resolved by 2-DE.

MATERIALS
Product Number
Brand
Product Description

Sigma-Aldrich
Tris(2-carboxyethyl)phosphine hydrochloride, BioUltra, ≥98% (NMR)