Skip to Content
Merck
CN
HomeSmall Molecules Analysis & QCAnalysis Atenolol and Chlorthalidone Tablets by TLC per USP

Identification of Atenolol and Chlorthalidone in Tablets by TLC acc. to USP and Quantification Using the TLC Explorer

Valanka D’Silva
R&D APAC Lab, Jigani, Bangalore, India

Abstract

This study presents a method for the identification of atenolol and chlorthalidone in tablet formulations using thin layer chromatography (TLC) as per the USP monograph. The TLC Explorer system was employed for enhanced digital documentation and automated evaluation, facilitating also quantitative analysis and by that extended the use of TLC for this application. Results indicate that the test samples exhibited Rf values comparable to standard solutions, confirming identity. Quantitative assessments yielded concentrations within 97.6% to 102.6% of nominal values, indicating the method's applicability. This application underscores the TLC Explorer's efficacy in pharmaceutical TLC analysis and documentation.

Section Overview

Introduction

Chemical structures of two pharmaceutical compounds, atenolol and chlorthalidone. On the left, the structure of atenolol includes a benzene ring attached to an amide group and an ether-linked side chain ending in a secondary amine with two methyl groups. It also has a hydroxyl group on the side chain. On the right, the structure of chlorthalidone features a fused ring system consisting of a benzothiadiazine ring with a chlorine atom, a sulfonamide group, and a ketone. A hydroxyl group is also attached to one of the rings. The structures are depicted in black lines with clear labeling of functional groups and atoms. The names “Atenolol” and “Chlorthalidone” are written below their respective structures.

Figure 1.Chemical structures of atenolol and chlorthalidone.

Atenolol and chlorthalidone (Figure 1) combination tablets are used to treat high blood pressure (hypertension). Atenolol (2-[p-[2-hydroxy-3-(isopropylamino)propoxy]phenyl]acetamide) is a cardio-selective beta blocker that works by affecting the response to some nerve impulses in the heart. As a result, the heart beats slower and decreases the blood pressure. Chlorthalidone (2-chloro-5-(1-hydroxy-3-oxoisoindolin-1-yl)benzene sulfonamide) is a diuretic that reduces the amount of water in the body by increasing the flow of urine, which helps to lower blood pressure.1

The United States Pharmacopeia (USP) monograph for atenolol and chlorthalidone tablets lists thin layer chromatography (TLC) as one of the methods for the identification test.2 In many pharmacopeial methods, TLC is a frequently cited technique for identity testing. HPTLC, a high-performance version of TLC and often used with automation, is a robust, reliable, rapid, and inexpensive tool used in qualitative and quantitative analysis of pharmaceutical compounds. This technique delivers chromatographic separations/fingerprints that can be visualized for identification/quantification and saved as electronic images for documentation.3-4

In this application note, the USP monograph specified identification test of atenolol and chlorthalidone in tablets by TLC is performed on the new TLC Explorer documentation system (Figure 2) that is also used to extend this method to a quantitative evaluation by video densitometry.

A compact laboratory instrument, TLC Explorer, with a white body and a glossy black front panel. The front features a rectangular screen in the upper center, likely used for displaying readings or navigation menus. To the right of the screen are three vertically aligned indicator labels marked “VIS,” “365,” and “254,” corresponding to different light wavelengths used for TLC plate visualization. Below the screen, the black surface includes a bright green and yellow Merck logo on the left and the brand name “Supelco” in yellow on the right. The left side of the device has a small protruding switch or button. The overall design is sleek and minimalistic, with the device sitting on a white surface, suggesting it is ready for benchtop use in an analytical laboratory setting.

Figure 2.TLC Explorer.

The TLC Explorer documentation system enables the digital and automated evaluation of TLC plates, enhancing the efficiency and accuracy of thin layer chromatography analysis. The device offers three illumination modes using LED light sources—white light (VIS), UV-A (366 nm), and UV-C (254 nm) – for the detection and fast measurement of the compounds of interest. The software offers special features like automated track recognition, simultaneous measurement of multiple plates and background signal correction. Overall, the TLC Explorer offers accurate TLC imaging for reliable densitometric measurements, enabling quantitative analysis and reliable data interpretation. 

Experimental

Reagent Preparation

Mobile phase: Dissolve 6.80 mL of ammonia solution 25% in 100 mL of water to obtain 1 N ammonium hydroxide. Mix 40 mL of this solution with 200 mL of n-butyl alcohol to obtain a mixture of n-butyl alcohol and 1 N ammonium hydroxide (5:1, v:v).

Standard Preparation

Atenolol

  • Standard stock solution 1 (50 mg/mL of atenolol): Weigh and dissolve 125 mg of atenolol in 2.5 mL of methanol.
  • Standard solution 1 (20 mg/mL atenolol): Dilute 400 µL of standard stock solution 1 to 1000 µL with methanol.

Chlorthalidone

  • Standard stock solution 2 (25 mg/mL chlorthalidone): Weigh and dissolve 62.5 mg of chlorthalidone in 2.5 mL of methanol.
  • Standard solution 2 (10 mg/mL chlorthalidone): Dilute 400 µL of standard stock solution 2 to 1000 µL with methanol.

Standard mix solutions I-VI

Prepare a total of six standard solutions (nos. I-VI) by pipetting 40 μL, 100 μL, 200 μL, 300 μL, 400 μL, and 440 μL each of standard stock solution 1 and standard stock solution 2 into six separate 2 mL vials. Add methanol to get to a final volume of 1 mL. The resulting solutions contain 2.0, 5.0, 10.0, 15.0, 20.0, and 22.0 mg/mL of atenolol and 1.0, 2.5, 5.0, 7.5, 10.0, and 11.0 mg/mL of chlorthalidone, respectively.

Sample Preparation

Sample tablets were used with an atenolol-to-chlorthalidone content ratio of 1:2.

  • Test solution I (for identification): Weigh and shake a quantity of powdered combination tablets, equivalent to 50 mg of atenolol (corresponding to 25 mg of chlorthalidone), with 2.5 mL of methanol for 15 minutes, and filter through a 0.45 µm PVDF membrane. Nominal resulting concentration of 20 mg/mL atenolol and 10 mg/mL chlorthalidone.
  • Test solution II (for quantification): Weigh and shake a quantity of powdered combination tablets, equivalent to 25 mg of atenolol (corresponding to 12.5 mg of chlorthalidone), with 2.5 mL of methanol for 15 minutes, and filter through a 0.45 µm PVDF membrane. Nominal resulting concentration of 10 mg/mL atenolol and 5 mg/mL chlorthalidone.
  • Recovery test solutions I-III: Prepare three samples by shaking a quantity of powdered tablets, equivalent to 25 mg of atenolol (corresponding to 12.5 mg of chlorthalidone), spiked with 125 μL, 250 μL, and 500 μL each of standard stock solution 1 and standard stock solution 2 for 15 minutes. Add methanol to get to a final volume of 2.5 mL and filter the solutions.

Instrument Parameters

Results

The identification of atenolol and chlorthalidone in tablets performed according to the USP monograph on the TLC Explorer under UV 254 nm and UV 366 nm is demonstrated in Figure 3. It additionally shows the calibration and recovery experiments performed on the TLC Explorer. Table 2 summarizes the obtained chromatographic results (Rf values). As required by the USP monograph for the identification, the principal spots obtained from the test solution (track 3) correspond to the Rf value size and intensity of the respective standard solutions (tracks 1 & 2). 

A thin-layer chromatography (TLC) plate viewed under 254 nm UV light, displaying two distinct horizontal bands labeled A and B. Band A, appearing as faint dark green spots at the lower part of the plate, corresponds to atenolol, while band B, appearing as sharp dark green or black spots higher on the plate, corresponds to chlorthalidone. Thirteen tracks are visible, marked numerically from 1 to 13 along the bottom edge in red, representing identification standards, calibration samples, a test sample, and recovery studies. The intensities of the bands vary across the tracks, with calibration tracks showing a gradient in spot intensity for both compounds.
A thin-layer chromatography (TLC) plate viewed under 366 nm UV light. Band A corresponding to atenolol is nearly invisible under this wavelength, while band B corresponding to chlorthalidone appears as bright white or bluish-white fluorescent spots near the top of each lane. Thirteen tracks are visible, marked numerically from 1 to 13 along the bottom edge in red, representing identification standards, calibration samples, a test sample, and recovery studies. The background is dark teal to black, enhancing the contrast of fluorescent spots. The overall pattern of band B confirms the successful identification, calibration, and recovery of chlorthalidone in the analyzed tablet samples.

Figure 3. TLC chromatogram demonstrating the identification studies (track 1-3), calibration studies (tracks 4-9), test sample (10) and recovery studies (tracks 11-13) of atenolol (A) and chlorthalidone (B) in tablets under UV 254 nm (left) and UV 366 nm (right) by the TLC Explorer.

Calibration

The results of the calibration experiments utilizing standard mix solutions I-VI and the calibration function of the TLC Explorer unit are displayed in Figures 4 & 5 Table 3 (for atenolol), and Figure 6 & 7 and Table 4 (for chlorthalidone).

A three-dimensional densitogram representing calibration studies for atenolol across six tracks. The graph plots intensity on the vertical axis, Rf values on the horizontal axis, and track numbers along the depth axis. Multiple colored lines—red, orange, green, blue, cyan, and magenta—each correspond to a different standard mix concentration. Two prominent sets of peaks are visible: one cluster appears at lower Rf values around 0.1 to 0.2, and another distinct set is observed at higher Rf values around 0.8, with increasing peak heights indicating higher analyte concentrations in successive tracks.
A two-dimensional densitogram showing the same calibration tracks plotted against Rf values on the x-axis and intensity on the y-axis. Here, the same-colored lines show two main sets of peaks corresponding to different compounds, with the higher Rf peaks being sharp and symmetrical, indicating consistent detection. The overlapping stacked nature of the lines demonstrates the increasing concentration gradient used for calibration.

Figure 4. 3D densitogram (left, tracks 4 to 9) and 2D densitogram (right) demonstrating the calibration studies for atenolol utilizing standard mix solutions I-VI.

A calibration curve plotting peak response on the vertical axis against concentration in micrograms per spot on the horizontal axis. Six circular data points, evenly distributed along the curve, correspond to increasing concentrations ranging from 20.0 to 220.0 micrograms per spot. A dotted trend line passes through the points, indicating a strong linear relationship. The linear regression equation y = 9.2898x + 132.12 is shown above the curve, along with the coefficient of determination, R² = 0.9994, signifying excellent correlation and indicating that the method provides highly consistent and reliable quantification of the analyte across the tested range.

Figure 5.Calibration curve obtained for atenolol using standard mix solutions I-VI
(c = 20.0, 50.0, 100.0, 150.0, 200.0 and 220.0 μg/spot).


A three-dimensional densitogram with intensity on the vertical axis, Rf values on the horizontal axis from left to right, and track numbers from back to front. Multiple colored lines represent individual tracks corresponding to different concentrations of chlorthalidone. Two prominent peaks are visible in each track at distinct Rf positions, indicating the presence of analyte bands. The peak height increases progressively from the nearest track to the farthest, illustrating calibration using ascending concentrations.
A two-dimensional densitogram graph with intensity plotted on the vertical axis and Rf values on the horizontal axis ranging from 0 to 1. Multiple overlapping curves, each corresponding to a different concentration level of chlorthalidone, show two distinct peaks. The first peak appears at a low Rf value around 0.1 and the second, more prominent peak is located around Rf 0.82. The peak heights increase systematically, reflecting higher concentrations in the standard solutions. Shaded areas beneath the curves emphasize the intensity differences between the bands. The background is white with light gridlines and black axis labels.

Figure 6. 3D densitogram (left, tracks 9 to 4) and 2D densitogram (right) demonstrating the calibration studies for atenolol utilizing standard mix solutions I-VI.

A calibration curve representing the linear relationship between concentration and peak response for chlorthalidone. The x-axis displays concentration in micrograms per spot, ranging from 0 to 120, while the y-axis shows peak response values ranging from 0 to 5600. Six individual data points are plotted as filled circles, and a dotted trendline passes through them, indicating a strong linear correlation. The equation of the line is given as y = 39.747x + 633.7, and the coefficient of determination (R²) is 0.9984, indicating excellent linearity. Both axis labels are bold, and the background is plain white.

Figure 7.Calibration curve obtained for chlorthalidone using standard mix solutions I-VI
(c = 10.0, 25.0, 50.0, 75.0, 100.0 and 110.0 μg/spot).

The determined sensitivities, LODs, and LOQs were 7.23 and 21.92 µg/spot for atenolol and 5.97 and 18.08 µg/spot for chlorthalidone, respectively. These represent concentrations in the applied sample solutions (10 µL) of 0.72 and 2.19 mg/mL for atenolol and of 0.59 and 1.80 µg/mL for chlorthalidone.

Recovery

The results of the spike recovery experiments utilizing recovery test solution I-III are displayed in Table 5 and Table 6 and were found to be in the range of 90.03 to 98.6%.

Results Test Sample

Quantification of the test sample II by applying the established calibration curve led to resulting concentrations of 97.64 µg/spot for atenolol and 51.32 µg/spot for chlorthalidone (Table 7), representing 97.64 % and 102.64 % of the respective nominal concentrations in the applied solutions (10 µL) of 10 mg/mL atenolol and 5 mg/mL chlorthalidone.

Conclusion

A method was applied following the USP monograph specified identification test of atenolol and chlorthalidone in tablets using TLC. The assessment and the documentation of the chromatographic results were done with the TLC Explorer documentation system. The principal spots in the chromatogram obtained from the test solution were similar in Rf value, size, and intensity to the principal spot in the chromatogram obtained with the standard solution as required by the monograph. Additionally, beyond the USP monographs TLC use for identification, the method was extended to a quantitative assessment for the two analytes using the quantification option of the TLC Explorer. The quantification of the test sample resulted in 97.6 -102.6% agreement with the nominal concentrations.

This application note demonstrates that the TLC Explorer documentation system serves as an efficient TLC visualizer, enabling data capture, track identification, and Rf value calculation.

Find more information on the TLC Explorer Documentation System.


Related Products

Device, Plates, Solvents & Reagents

Loading

Reference Materials

Loading

References

1.
Youssef RM, Maher HM, El-Kimary EI, Hassan EM, Barary MH. 2013. Validated Stability-Indicating Methods for the Simultaneous Determination of Amiloride Hydrochloride, Atenolol, and Chlorthalidone Using HPTLC and HPLC with Photodiode Array Detector. Journal of AOAC International. 96(2):313-323. https://doi.org/10.5740/jaoacint.11-347
2.
USP monograph: Atenolol and Chlorthalidone tablets. USP–NF. https://doi.usp.org/USPNF/USPNF_M6338_01_01
3.
Attimarad M, Mueen Ahmed K, Aldhubaib BE, Harsha S. 2011. High-performance thin layer chromatography: A powerful analytical technique in pharmaceutical drug discovery. Pharmaceutical Methods. 2(2):71-75. https://doi.org/10.4103/2229-4708.84436
4.
Srivastava M. 2011. An Overview of HPTLC: A Modern Analytical Technique with Excellent Potential for Automation, Optimization, Hyphenation, and Multidimensional Applications. Springer.3-24. https://doi.org/10.1007/978-3-642-14025-9_1
Sign In To Continue

To continue reading please sign in or create an account.

Don't Have An Account?