产品名称
Opticap XL 600 Millipore Express SHF 可γ辐照 0.2 µm 3/4 in.TC-9/16 in.HB, Opticap® XL, inlet connection diam. 3/4 in., pore size 0.2 μm, cartridge nominal length 6.5 in. (16.5 cm)
material
polyester support
polyethersulfone membrane
polypropylene housing
polypropylene vent cap
silicone O-ring
silicone seal
reg. compliance
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
sterility
non-sterile
sterilization compatibility
autoclavable compatible
gamma compatible
x-ray compatible
product line
Opticap® XL
feature
hydrophilic
manufacturer/tradename
Millipore Express®
parameter
1 bar max. differential pressure (15 psi) at 80 °C
1 bar max. inlet pressure (15 psi) at 80 °C
2.1 bar max. differential pressure (30 psi) at 25 °C (Reverse; Intermittent)
2.8 bar max. inlet pressure (40 psi) at 60 °C
5.5 bar max. differential pressure (80 psi) at 25 °C
5.5 bar max. inlet pressure (80 psi) at 25 °C
6.9 bar max. differential pressure (100 psi) at 25 °C (Forward; Intermittent)
6.9 bar max. inlet pressure (100 psi) at 25 °C (Intermittent)
technique(s)
sterile filtration: suitable
L
16.5 cm (6.5 in.)
W
2.2 in.
cartridge nominal length
6.5 in. (16.5 cm)
diam.
5.6 cm (2.2 in.)
filtration area
0.097 m2 (1.046 ft2)
inlet connection diam.
3/4 in.
inlet to outlet W
16.5 cm (6.5 in.)
outlet connection diam.
9/16 in.
impurities
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
matrix
Millipore Express® SHF
pore size
0.2 μm pore size
bubble point
≥4000 mbar (58 psig), air with water at 23 °C
fitting
19 mm (3/4 in.) inlet sanitary flange
14 mm (9/16 in.) outlet hose barb
Quality Level
Analysis Note
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
After sterilization and a controlled water flush of 3.0 L, samples exhibited less than 500 ppb TOC per USP <64> and less than 1.3 µS/cm per USP <64> at 25 °C.
General description
Disclaimer
Other Notes
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer installation section of OPTICAP® XL Capsules and OPTICAP® XLT Capsules user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Packaging
Preparation Note
Gamma compatible to 45 kGy; 3 autoclave cycles of 60 min @ 123 °C; not in-line steam sterilizable
Legal Information