ASTM D6526: GC Analysis of Impurities in Toluene on SLB®-IL100, 30 m Column

ASTM D6526: GC Analysis of Impurities in Toluene on SLB®-IL100, 30 m Column suitable for GC

Materials

analytical column

Product No.
Description
Pricing

SLB®-IL100 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

SLB-IL100, 30 m x 0.25 mm I.D., 0.20 μm (28884-U)

oven

70 °C

inj. temp.

200 °C

detector

FID, 200 °C

carrier gas

helium, 28 cm/sec

injection

1 μL 40:1 split

liner

4 mm I.D., split type, cup design

sample

toluene, with each aromatic impurity spiked at 0.01% by weight