Skip to Content
Merck
CN

GC Analysis of a QC Test Mix on SLB®-ILPAH, 0.10 mm I.D. Column

GC Analysis of a QC Test Mix on SLB®-ILPAH, 0.10 mm I.D. Column suitable for GC

CONDITIONS

column

SLB-ILPAH, 15 m x 0.10 mm I.D., 0.03 μm (29798-U)

oven

90 °C (6 min), 20 °C/min to 225 °C, 5 °C/min to 300 °C (10 min)

inj. temp.

300 °C

detector

FID, 310 °C

carrier gas

hydrogen, 0.8 mL/min, constant flow

injection

1 μL, 500:1 split

liner

2.3 mm I.D., split/splitless type, wool packed straight FocusLiner design

sample

10 PAHs, each at 100 μg/mL in methylene chloride

Description

Analysis Note

PAHs are present in oil, coal, and tar deposits. They can also be produced by the incomplete combustion of organic matter or when grilling meat. Monitoring is important because they are identified as carcinogens. Multiple isomers exist, which are difficult to resolve chromatographically. This application depicts the QC test mix analyzed on a SLB ILPAH column. As shown, this column has the necessary selectivity to resolve several isomer sets.

Legal Information

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany