GC Analysis of C18:0, C18:1, C18:2, and C18:3 FAME Isomers on SLB®-IL111

Materials
related product
Product No.
Description
Pricing
CONDITIONS
column
SLB-IL111, 200 m x 0.25 mm I.D., 0.20 μm (29689-U)
oven
150 °C
inj. temp.
250 °C
detector
FID, 250 °C
carrier gas
hydrogen
injection
1μL, 50:1 split
liner
Split type, cup design
sample
(Mix of C18:0, C18:1 (from partially hydrogenated vegetable oil [PHVO]), C18:2, and C18:3 FAME isomers)
Description
Analysis Note
Partially hydrogenated vegetable oil, linoleic acid, and linolenic acid FAMEs analyzed on the 200 meter SLB-IL111 capillary GC column.
Legal Information
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany