Skip to Content
Merck
CN

GC Analysis of Class 2 Residual Solvents on VOCOL® after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Class 2 Residual Solvents on VOCOL® after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Materials

used together

Product No.
Description
Pricing

SPME Fiber Assembly

85 μm Polyacrylate, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, white hub

CONDITIONS

sample/matrix

2 mL water + sodium chloride at 25% (pH 11) containing each analyte at 5 ppm, in a 4 mL vial

SPME fiber

85 μm Polyacrylate (57304)

extraction

heated headspace, 60 °C for 5 min

desorption process

250 °C for 3 min

column

VOCOL, 10 m x 0.20 mm I.D., 1.2 μm (24129-U)

oven

40 °C (0.75 min), 20 °C/min to 200 °C (10 min)

inj. temp.

250 °C

detector

FID

carrier gas

helium, 35 cm/sec @ 40 °C

injection

5:1 split

liner

0.75 mm I.D. SPME type

Description

Legal Information

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany