GC Analysis of Volatiles in Water on Equity-1 after SPME using 30 μm PDMS Fiber

GC Analysis of Volatiles in Water on Equity™-1 after SPME using 30 μm PDMS Fiber suitable for GC, application for SPME

Materials

SPME fiber

Product No.
Description
Pricing

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 30 μm(PDMS, needle size 24 ga, for use with autosampler

CONDITIONS

sample/matrix

0.7 mL water spiked with 2-600 μg/liter each analyte, 0.25 g NaCl added

SPME fiber

30 μm PDMS (57309)

extraction

headspace, 12 sec (no stirring), ambient temp.

desorption process

2-3 min, 250 °C (0.75-1.0 mm I.D. inj. port liner)

column

Equity-1, 10 m × 0.20 mm I.D., 1.2 μm (available on request)

oven

70 °C (0.2 min) to 180 °C at 50 °C/min

detector

FID, 250 °C

Description

Legal Information