GC Analysis of Water (0.1-100% Calibration Curve) on Watercol 1460

GC Analysis of Water (0.1-100% Calibration Curve) on Watercol™ 1460 suitable for GC

CONDITIONS

column

Watercol 1460, 30 m x 0.25 mm I.D., 0.20 μm (29695-U)

oven

60 °C (12 min)

inj. temp.

250 °C

detector

TCD, 200 °C

carrier gas

helium, 26 cm/sec

injection

0.5 μL for 0.1-60% water, 0.1 μL for 80-100% water, 100:1 split

liner

4 mm I.D., split type, cup design

sample

7 standards, water at 0.1, 0.5, 20, 40, 60, and 80% (v/v), each in ethanol, and water at 100%

Description

Analysis Note

The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water can be integrated and quantified. This data shows the linearity obtained from a 7-point calibration curve (0.1-100%) for water on Watercol 1460.