US EPA Method 624: GC Analysis of Volatiles on VOCOL® (30 m x 0.53 mm I.D.) after Purge & Trap using "K" Trap

Materials
related product
Product No.
Description
Pricing
CONDITIONS
sample/matrix
20 ppb each component in 5 mL water (25320-U)
purge trap
VOCARB 3000 (21066-U)
purge
40 mL/min for 11 min
dry purge
3 min
desorption temp.
250 °C for 4 min
bake
280 °C for 10 min
column
VOCOL®
, 30 m × 0.53 mm I.D., 3.0 μm (25320-U)
oven
5 °C (2 min), 5 °C/min to 200 °C
scan range
m/z 35-260 at 0.6 sec/scan
carrier gas
helium, 7.5 mL/min
Description
Legal Information
VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany