Merck
CN

US EPA Method 8015 (modified): GC Analysis of Diesel Range Organics (DRO) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm), Fast GC Analysis

US EPA Method 8015 (modified): GC Analysis of Diesel Range Organics (DRO) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm), Fast GC Analysis suitable for GC

Materials

analytical column

Product No.
Description
Pricing

SPB®-608 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SPB®-608 Capillary GC Column

L × I.D. 30 m × 0.53 mm, df 0.50

Equity-1701 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

Equity-1701 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.00 μm

Equity-1701 Capillary GC Column

L × I.D. 30 m × 0.53 mm, df 0.50 μm

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.00 μm

SLB®-5ms Capillary GC Column

L × I.D. 15 m × 0.10 mm, df 0.10 μm

SLB®-5ms Capillary GC Column

L × I.D. 20 m × 0.18 mm, df 0.18 μm

accessory

Product No.
Description
Pricing

Wheaton wide-mouth bottle

capacity 32 oz (1000 mL), white polypropylene cap, PTFE flat liner, amber

Wheaton straight-mouth bottles with cap, clear

white polypropylene cap, capacity 8 oz (250 mL), PTFE flat liner

guard column

Product No.
Description
Pricing

Intermediate Polar Fused Silica Tubing

L × I.D. 5.0 m × 0.25 mm

Intermediate Polar Fused Silica Tubing

L × I.D. 5.0 m × 0.53 mm

recommended

Product No.
Description
Pricing

Polyimide Sealing Resin

pkg of 5 g

required but not provided

Product No.
Description
Pricing

"Y" GlasSeal Connector, Borosilicate Glass

pkg of 1 ea

"Y" GlasSeal Connector, Fused Silica

pkg of 1 ea

"Y" GlasSeal Connector, Fused Silica

pkg of 3 ea

standard

Product No.
Description
Pricing

Underground Storage Tank (UST) Modified DRO

certified reference material, TraceCERT®, 1000 μg/mL each component in hexane, ampule of 1 mL

CONDITIONS

column

SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)

oven

50 °C (2 min), 20 °C/min to 325 °C (1 min)

inj. temp.

250 °C

detector

FID, 350 °C

carrier gas

helium, 0.9 mL/min, constant

injection

1 μL, splitless (splitter open at 0.5 min)

liner

2 mm I.D., FocusLiner inlet liner

sample

each analyte at 50 μg/mL in methylene chloride

Description

Legal Information

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany