US EPA Method 8082: GC Analysis of Aroclor 1016/1260 (PCBs) on SLB®-5ms (15 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

Materials
accessory
Product No.
Description
Pricing
Wheaton wide-mouth bottle
amber soda-lime glass bottle, capacity 1000 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 earelated product
Product No.
Description
Pricing
Pesticide Surrogate Spike Mix
certified reference material, TraceCERT®, 200 μg/mL each component in acetone, ampule of 1 mLused together
Product No.
Description
Pricing
CONDITIONS
column
SLB-5ms, 15 m × 0.10 mm I.D., 0.10 μm (28466-U)
oven
80 °C (0.5 min), 50 °C/min to 200 °C, 35 °C/min to 360 °C (2 min)
inj. temp.
225 °C
detector
ECD, 360 °C
carrier gas
hydrogen, 40 cm/sec constant
injection
2 μL, splitless (0.75 min)
liner
4 mm I.D., single taper
sample
Aroclor standard mix 1 (46846-U) diluted to 500 ppb / 50 ppb (Aroclors / surrogates) in n-hexane
Description
Legal Information
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany