Aniracetam Capsules Assay (HPLC) following Chinese Pharmacopeia 2015 Monograph Method on Ascentis® C18 with UV Detection

Aniracetam Capsules Assay (HPLC) following Chinese Pharmacopeia 2015 Monograph Method on Ascentis® C18 with UV Detection application for HPLC

Materials

related product

Product No.
Description
Pricing

Ascentis® C18 HPLC Column

5 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

column

Ascentis® C18, 15 cm x 4.6 mm, 5 μm particles (581324-U)

mobile phase

[A} water; [B] acetonitrile, (65:35, A:B)

flow rate

1.0 mL/min

pressure

1305 psi (90 bar)

column temp.

25 °C

injection

10 μL

sample

SST Solution

Description

Analysis Note

LC-UV Analysis of Aniracetam Capsules (Huilaxitan Jiaonang). Aniracetam, or N-anisoyl-2-pyrrolidinone, is a nootropic agent (substance that improve cognitive function). It is as a prescription drug world-wide, but not approved for use in the United States. The performance criteria per the Chinese pharmacopeia (ChP) 2015 monograph method are met. The detection limit (LOD) and limit of quantitation (LOQ) with HPLC-UV are both better than 1 ppm.

Other Notes

SST Solution: Dissolve 50 mg of Aniracetam with 5 mL methanol into colorimeter tube, heat in water bath at 70 °C for 1h. Cool down to room temperature. Dilute with mobile phase to 1 mg/mL solution.
Standard solution: Dissolve appropriate amount of Aniracetam to mobile phase to obtain 80 μg/mL (80 ppm) standard solution.
Test solution: Take appropriate amount of drug powder from capsules into mobile phase to obtain 80 μg/mL (80 ppm) sample solution. Filter sample solution prior to injection through a 0.45 μm Millex PVDF filter.