HPLC Analysis of Carbidopa on Chromolith® High Resolution RP-18e

CONDITIONS
column
Chromolith® High Resolution RP-18 endcapped, 100 x 4.6 mm (1.52022.0001)
mobile phase
[A] methanol; [B] water + 0,1%TFA;
gradient
100% B held for 0.01 min; to 80% B in 0.9 min; held at 80% A for 9 min (Original: Initial composition: 0-0.1min 100%B, 1-10min 100%B to 80%B)
flow rate
2.5 mL/min
column temp.
ambient
detector
282 nm
injection
1 μL
sample
2,4,5-Trihydroxyphenylalanine 4μg/ml, Levodopa 402μg/ml, Methyldopa 384μg/ml, Daopamine 217μg/ml, Carbidopa 232μg/ml, 3,4-Dihydroxyphenyl acetic acid 4μg/ml, 3-o-Methylcarbidopa 4μg/ml
Description
Legal Information