Skip to Content
Merck
CN

HPLC Analysis of DNPH Carbonyl Compounds on Ascentis® Express C18

HPLC Analysis of DNPH Carbonyl Compounds on Ascentis® Express C18 application for air monitoring, application for HPLC

Materials

related product

Product No.
Description
Pricing

ATIS Adsorbent Tube Injector System 110V

ATIS Purge and Trap/Humidifier Module

ATIS Extraction Glassware with Micro Connector

Acetone-2,4-DNPH

analytical standard

Formaldehyde-2,4-DNPH

analytical standard

Propionaldehyde-2,4-DNPH

analytical standard

Valeraldehyde-2,4-DNPH

analytical standard

Ascentis® Express C18 (2.7 μm) HPLC Column

L × I.D. 15 cm × 4.6 mm, pk of 1 ea

used together

Product No.
Description
Pricing

LpDNPH S10 Cartridge

bed wt. 350 mg, volume 3 mL, pkg of 50 ea

CONDITIONS

sample/matrix

Wood Composite/Air

SPE tube/cartridge

Pk 50 LpDNPH S10 350 mg/3 mL, BULK (23124-U)

elution

1.0 mL acetonitrile

column

Ascentis Express C18, 15 cm x 4.6 mm, 2.7 μm particles (53829-U)

gradient

65%A held for 1 min; to 67% A in 0.5 min; held for 6.5 min; to 70% A in 3min; held for 6 min

flow rate

0.8 mL/min

column temp.

24 °C

detector

UV, 360 nm

injection

10 μL