HPLC Analysis of E. coli lysate with various pore size Zenix® columns

HPLC Analysis of E. coli lysate with various pore size Zenix® columns application for HPLC

Materials

related product

Product No.
Description
Pricing

Zenix® SEC-150 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 7.8 mm, 3 μm particle size, stainless steel

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 7.8 mm, 3 μm particle size, stainless steel

CONDITIONS

column

Zenix SEC-150, 30 cm x 7.8 mm I.D., 3 μm particles (Z777018)

column

Zenix SEC-300, 30 cm x 7.8 mm I.D., 3 μm particles (Z777033)

mobile phase

150 mM Phosphate Buffer, pH 7

flow rate

1.0 mL/min

column temp.

Ambient

detector

UV, 214nm

injection

10 μL

sample

E. coli lysate, 2.5 mg/mL

Description

Analysis Note

An cell lysate, produced by destroying the membranes of E. Coli cells, is analyzed on two Zenix gel filtration columns packed with 3 micron particles. The top chromatogram shows the peak profile on a 150 Å pore size Zenix SEC-150 gel filtration column, while the bottom chromatogram shows the same sample on a 300 Å pore size Zenix SEC-300 column. As expected, the peaks shift to longer retention times on the SEC-300 column. Although roughly the same number of components are visible in each chromatogram, the SEC-150 column does a better job of pulling the peaks apart, which helps when fractions need to be collected for further analysis.