Skip to Content
Merck
CN

HPLC Analysis of Organic Acids in Beverages on Chromolith® HighResolution RP-18 endcapped

HPLC Analysis of Organic Acids in Beverages on Chromolith® HighResolution RP-18 endcapped application for HPLC

Materials

related product

Product No.
Description
Pricing

Citric acid

99%

used together

Product No.
Description
Pricing

Ammonium phosphate dibasic

≥99.99% trace metals basis

CONDITIONS

sample preparation

sonicate 5 mL of orange drink for 5 minutes, add 0.2 mL of phosphoric acid solution (1M), dilute to 10 mL with water

column

Chromolith HighResolution RP-18, 10 cm x 4.6 mm I.D., 2 μm particles (1.52022)

mobile phase

10 mM diammonium hydrogen phosphate, pH 2.7

flow rate

1.0 mL/min

pressure

32 bar (464 psi)

column temp.

30 °C

detector

UV, 210 nm

injection

20 μL

Description

Legal Information