HPLC Analysis of Poly(ethyleneglycol) polymers on SRT® SEC-150 and SRT® SEC-300

HPLC Analysis of Poly(ethyleneglycol) polymers on SRT® SEC-150 and SRT® SEC-300 application for HPLC

Materials

related product

Product No.
Description
Pricing

SRT® SEC-150 Gel Filtration Column, 5 μm

L × I.D. 30 cm × 7.8 mm, 5 μm particle size, stainless steel

SRT® SEC-300 Gel Filtration Column, 5 μm

L × I.D. 30 cm × 7.8 mm, 5 μm particle size, stainless steel

used together

Product No.
Description
Pricing

SRT® SEC-150 Gel Filtration Guard Column, 5 μm

L × I.D. 5 cm × 7.8 mm, 5 μm particle size, stainless steel

SRT® SEC-300 Gel Filtration Guard Column, 5 μm

L × I.D. 5 cm × 7.8 mm, 5 μm particle size, stainless steel

CONDITIONS

column

SRT-150, 30 cm x 7.8 mm I.D., 5 μm particles, 150 Å (Z777045)

column

SRT-300, 30 cm x 7.8 mm I.D., 5 μm particles, 150 Å (Z777051)

mobile phase

150 mM phosphate Buffer, pH 7.0

flow rate

1.0 mL/min

pressure

720 psi (49.6 bar)

detector

RI(30 C°)

injection

20 μL

sample

1.5 mg/mL in mobile phase PEG 1. MW 636, 6,550, 118,000, 932,000; PEG 2. MW 2010, 42,700, 496,000 ; PEG 3. MW 232, 17,900, 271,000

Description

Analysis Note

Although proteins are the most common applications of Sepax SRT and Zenix gel filtration columns, they can also be used to other water-soluble polymers such as the poly(ethyleneglycol) samples shown in these chromatograms. The conditions are very similar to those used for protein analysis, namely 150mM phosphate buffer at pH 7.0. Judging by the narrow peak widths all PEG polymers shwon here have rather narrow size distributions.