HPLC Analysis of the Monoclonal Antibody (mAb) Erbitux (Cetuximab) on BIOshell A400 Protein C4

HPLC Analysis of the Monoclonal Antibody (mAb) Erbitux (Cetuximab) on BIOshell™ A400 Protein C4 application for HPLC

Materials

related product

Product No.
Description
Pricing

BIOshell A400 Protein C4, 3.4 μm HPLC Column

3.4 μm particle size, L × I.D. 10 cm × 2.1 mm

Guard Cartridge Holder for BIOshell Guard Column

used together

Product No.
Description
Pricing

BIOshell A400 Protein C4, 3.4 μm Guard Cartridge

3.4 μm particle size, L × I.D. 0.5 cm × 2.1 mm, pkg of 3 ea

CONDITIONS

column

BIOshell A400 Protein C4, 10 cm X 2.1 mm I.D., 3.4 μm particles (66825-U)

mobile phase

(A) 80:20, 0.1% TFA in water: 0.1% TFA in acetonitrile; (B) 20:80, 0.1% TFA in water: 0.1% TFA in acetonitrile

gradient

20 to 35%B in 10 min (or 5 min)

flow rate

0.2 mL/min

pressure

470 psi (or 940 psi)

column temp.

90 °C

detector

UV, 215 nm

injection

0.4 μL

sample

1 g/L mAb in 20 mM Tricine/KOH, pH 7

Description

Analysis Note

Antibodies can exhibit considerable hydrophobicity, and so efficient reversed-phase elution of this intact mAb requires the use of high temperature. The top chromatogram illustrates that BIOshell Fused-Core® columns can be operated at high flow rate while maintaining good peak shape at a reasonable back pressure.