CONDITIONS

column

Ascentis Express Phenyl-Hexyl, 10 cm x 2.1 mm I.D., 2.7 μm particles (53336-U)

gradient

[A] 10 mM ammonium acetate; [B] acetonitrile

gradient

2% B for 1 min; to 100% B in 10 min; held at 100% B for 1 min

flow rate

0.3 mL/min

pressure

2930 psi (202 bar)

column temp.

35 °C

detector

MS, ESI+, XIC

injection

0.5 μL

sample

Antifungals, 1-10 μg/mL in 99:1, water:methanol