Merck
CN

LC/MS (TOF) Analysis of Warfarin Enantiomers in Plasma on Astec® CHIROBIOTIC® V after SPE using HybridSPE®-Phospholipid

LC/MS (TOF) Analysis of Warfarin Enantiomers in Plasma on Astec® CHIROBIOTIC® V after SPE using HybridSPE®-Phospholipid application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Astec® CHIROBIOTIC® V Chiral HPLC Column

5 μm particle size, L × I.D. 10 cm × 4.6 mm

SPE tube or plate

Product No.
Description
Pricing

HybridSPE®-Phospholipid

96-well Plate, bed wt. 50 mg, volume 2 mL, pk of 1

CONDITIONS

sample/matrix

Rat plasma stabilized with K2EDTA was acquired from Lampire Biological Laboratories, (Pipersville PA). (Plasma was spiked directly from stock standard to a level of 100 ng/mL in 3:1 (plasma:1% formic acid acetonitrile).)

SPE tube/cartridge

Pk1 50 mg/96-Well HybridSPE-Phospholipid Plate (575656-U)

condition

Apply 100 μL of plasma to plate, followed by 300 μL of 1% formic acid acetonitrile. Agitate via vortex for 4 min.

elution

Place on vacuum manifold and apply 10” Hg vacuum for 4 minutes. Collect filtrate and analyze directly.

column

CHIROBIOTIC V,10 cm x 4.6 mm I.D., 5 μm particles (11022AST)

mobile phase

[A] 0.1% formic acid water; [B] acetonitrile, (75:25, A:B)

flow rate

1 mL/min

column temp.

35 °C

detector

ESI+, 100-1000 m/z

injection

2 μL

Description

Analysis Note

The R- and S- forms of warfarin, a commonly prescribed anticoagulant, are cleared by the body via different pathways and have different pharmacokinetic and pharmacodynamic effects. An Astec CHIROBIOTIC V HPLC column provided rapid resolution of the enantiomers under conditions that are compatible with mass spectrometric (MS) detection. By using HybridSPE-Phospholipid for the sample prep, ion-suppressing phospholipids were removed for improved sensitivity and longer column lifetime. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant reference material grade standards provided reliable quantification.

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC
HybridSPE is a registered trademark of Merck KGaA, Darmstadt, Germany
null