Skip to Content
Merck
CN

HPLC Analysis of β-Blockers and β-Agonists in Urine and Wastewater on a C18 Column after SPE using SupelMIP® SPE-Beta-Receptor

HPLC Analysis of β-Blockers and β-Agonists in Urine and Wastewater on a C18 Column after SPE using SupelMIP® SPE-Beta-Receptor application for SPE, application for LC-MS

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

horse urine or waste water spiked with 10 beta-agonists and beta-blockers at 1 ng/mL

SPE tube/cartridge

SupelMIP SPE - Beta-Receptor, 25 mg/10 mL (LRC) (53223-U)

condition

1 mL acetonitrile; 1 mL DI water

sample addition

1 mL

washing

3 x 1 mL DI water; 1 mL acetonitrile; 1 mL 60% acetonitrile/40% DI water; apply 2 min of full vacuum to dry the tube after each step

elution

2 x 1 mL 1% formic acid in acetonitrile

column

C18, 5 cm x 3 mm I.D., 3 μm particles

eluate post-treatment

evaporate under nitrogen and reconstitute with 150 μL 5% acetonitrile in 10 mM ammonium acetate, pH 4.6 prior to LC-MS-MS analysis

mobile phase

[A] 10 mM ammonium acetate, pH 4.6 (adjusted with acetic acid); and [B] acetonitrile

flow rate

0.5 mL/min

column temp.

ambient

detector

MS/MS, ESI(+)

injection

20 μL

Description

Application

Sample Pre-treatment:
Horse urine was centrifuged at 3000 g for 10 min, diluted 1:1 (v/v) with DI water, adjusted to pH 7. Wastewater was filtered with 1 μm filter paper and adjusted to pH 6-7. Samples were spiked with 10 beta-agonists and beta-blockers at the level of 1 ng/mL.