LC/MS Analysis of Ethanol Metabolites in Urine on Ascentis® Express OH5 after Dilution with Deuterated Internal Standards

LC/MS Analysis of Ethanol Metabolites in Urine on Ascentis® Express OH5 after Dilution with Deuterated Internal Standards application for LC-MS

Materials

related product

Product No.
Description
Pricing

Ascentis® Express OH5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

Ethyl-β-D-glucuronide

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Ethyl-β-D-glucuronide-(ethyl-d5)

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Ethyl sulfate sodium salt

1.0 mg/mL in methanol (as ethyl sulfate), ampule of 1 mL, certified reference material, Cerilliant®

Ethyl-d5 sulfate sodium salt

1.0 mg/mL in methanol (as ethyl sulfate), ampule of 1 mL, certified reference material, Cerilliant®

used together

Product No.
Description
Pricing

Ammonium formate

eluent additive for LC-MS, LiChropur, ≥99.0%

CONDITIONS

column

Ascentis Express OH5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53757-U)

mobile phase

[A] 5 mM ammonium formate and 0.1% formic acid in 95:5, acetonitrile:water; [B] 5 mM ammonium formate and 0.1% formic acid in 80:20, acetonitrile:water

gradient

0 to 100% B in 1 min, held at 100% B for 4 min, to 0% B, held for 5 min

flow rate

0.4 mL/min

pressure

1450 psi (100 bar)

column temp.

30 °C

detector

MS, ESI(-), 125.0 m/z, 126.1 m/z, 221.1 m/z, 226.2 m/z

injection

5 μL

sample

urine spiked with ethyl sulfate and ethyl-β-D-glucuronide, both at 50 ng/mL, diluted 20:1 in acetonitrile containing deuterated internal standards at 100 ng/mL

Description

General description

Ethyl sulphate (EtS) and ethyl glucuronide (EtG) are direct ethanol metabolites and may indicate recent alcohol consumption. The two compounds differ in their pathways for formation and degradation. Being polar compounds, they are poorly retained by C18 phases and elute early in the chromatogram along with matrix compounds. This leads to poor or unreliable quantification. The method shown here uses HILIC mode on an Ascentis Express OH5 column to retain both analytes well, making it very likely to be more robust and reliable, as well as highly MS-friendly. Cerilliant CRMs provided reliable quantification. The internal standard was needed for accurate quantification. External calibration resulted in a large excess of sulfate.